FDA Medical Device Recalls (openFDA)
Physio Control, Inc.: Internal Defibrillation Electrodes: PN: 805355-10 thru 14 and 805355-020 thru -024.
- Recall number
- Z-2036-2008
- Recalling firm
- Physio Control, Inc.
- Product
- Internal Defibrillation Electrodes: PN: 805355-10 thru 14 and 805355-020 thru -024.
- Status
- Terminated
- Initiated
- 2008-01-29
- Posted
- 2008-09-16
- Terminated
- 2011-05-04
- Reason
- Based on testing, sterilization methods described in labeling may be ineffective and may cause damage or corrosion to the paddles or handles. If handles and/or paddles are not sterilized adequately and are used on a patient, there is a risk of infection. If handle discharge button is damaged during sterilization, defibrillation therapy may be prevented.
- Root cause
- Labeling design
- Action
- Starting 2/5/08, Physio-Control sent letters to customers which notifies them not to use Flash Gravity Steam, 12/88 Ethylene Oxide, STERRAD 100 machine, or Gravity Steam to sterilize. The firm recommends the use of 100% EO, STERRAD 100S machine, or Prevacuum steam for sterilization. The letter also recommends that after sterilization, the handles and paddles should be examined carefully for damage. If damaged, the product should be removed from use. The firm will follow up with a revised Sterilization Guideline document and requests posting the letter in the sterilization department in the meantime. Contact technical support at 800-442-1142 if you have questions. On 7/23/08 the firm issued new sterilization guidelines that supersede all others.
- Quantity
- not on file
- Distribution
- Distributed to hospitals and medical centers. Nationwide, DOD, and international: ARGENTINA, BAHRAIN, BANGLADESH, BERMUDA, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, INDIA, IRAN, ISRAEL, JAPAN, KOREA, LEBANON, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, QATAR, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SRI LANKA, SYRIA, THAILAND, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIRGIN ISLANDS - US, YEMEN, ZIMBABWE
- Lot, serial or model codes
- Part Number 805355 - Lot Numbers: 0001, 0060, 0080, 0090, 0100, 0102, 0104, 0106, 0107, 0123, 0197, 0200, 0203, 0204, 0205, 0206, 0207, 0300, 0301, 0304, 0305, 0306, 0307, 0398, 0399, 0404, 0405, 0407, 0499, 0500, 0501, 0502, 0503, 0504, 0505, 0507, 0600, 0601, 0604, 0605, 0606, 0607, 0698, 0700, 0703, 0704, 0705, 0706, 0707, 0800, 0802, 0803, 0804, 0805, 0806, 0807, 0895, 0899, 0900, 0901, 0903, 0904, 0905, 0906, 0907, 1000, 1001, 1002, 1004, 1005, 1006, 1007, 1023, 1097, 1099, 1101, 1103, 1104, 1105, 1106, 1107, 1201, 1202, 1204, 1205, 1206, 1207, 1298, 1300, 1302, 1304, 1305, 1306, 1307, 1398, 1400, 1402, 1403, 1404, 1405, 1406, 1407, 1498, 1499, 1501, 1502, 1505, 1506, 1601, 1602, 1603, 1604, 1605, 1606, 1607, 1700, 1701, 1702, 1703, 1704, 1705, 1706, 1707, 1797, 1799, 1801, 1802, 1804, 1805, 1806, 1807, 1898, 1899, 1903, 1905, 1906, 1907, 1999, 2000, 2001, 2002, 2003, 2004, 2005, 2006, 2101, 2103, 2104, 2105, 2106, 2107, 2202, 2203, 2205, 2206, 2207, 2298, 2299, 2300, 2301, 2302, 2304, 2305, 2306, 2399, 2402, 2403, 2405, 2406, 2407, 2499, 2500, 2501, 2502, 2503, 2505, 2506, 2507, 2599, 2600, 2601, 2602, 2603, 2605, 2606, 2607, 2698, 2699, 2702, 2800, 2802, 2806, 2807, 2899, 2900, 2901, 2903, 2904, 2905, 2906, 2907, 2999, 3000, 3001, 3002, 3005, 3006, 3007, 3098, 3099, 3100, 3101, 3102, 3104, 3105, 3106, 3107, 3197, 3198, 3203, 3204, 3205, 3206, 03294, 3299, 3302, 3303, 3305, 3306, 3398, 3399, 3400, 3401, 3403, 3404, 3405, 3406, 3407, 3498, 3499, 3500, 3501, 3503, 3505, 3507, 3597, 3599, 3602, 3603, 3605, 3606, 3607, 3697, 3698, 3699, 3700, 3701, 3702, 3705, 3706, 3707, 3798, 3799, 3805, 3806, 3807, 3899, 3900, 3901, 3905, 3906, 3907, 3998, 4000, 4003, 4005, 4006, 4007, 4099, 4100, 4101, 4102, 4103, 4105, 4106, 4107, 4197, 4199, 4200, 4201, 4203, 4204, 4206, 4207, 4298, 4299, 4300, 4302, 4304, 4305, 4306, 4307, 4399, 4401, 4405, 4406, 4498, 4499, 4500, 4502, 4503, 4506, 4507, 4600, 4601, 4603, 4605, 4606, 4607, 4699, 4701, 4702, 4704, 4705, 4706, 4797, 4799, 4803, 4805, 4806, 4900, 4904, 4906, 4907, 4998, 5000, 5001, 5002, 5003, 5005, 5006, 5103, 5105, 5106, 5199, 5200, 5203, 5204, 5205, 5206, 5299, 5301.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LDD
- Firm FEI
- 3015876
- Firm city
- Redmond
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28