Skip to the content

FDA Medical Device Recalls (openFDA)

Joerns Healthcare: ProCair Mattress: (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 (c) PROCAIR PLUS 84" MATTRESS, MATERIAL NUMBER 222-6384 Product Usage: Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).

Recall number
Z-2034-2018
Recalling firm
Joerns Healthcare
Product
ProCair Mattress: (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 (c) PROCAIR PLUS 84" MATTRESS, MATERIAL NUMBER 222-6384 Product Usage: Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).
Status
Terminated
Initiated
2018-05-03
Posted
not on file
Terminated
2023-09-28
Reason
Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.
Root cause
Employee error
Action
The firm initiated the recall by letter on 05/04/2018. The letter requested the consignee identify potentially affected mattresses and contact Joerns Healthcare directly to schedule on-site rework of the unit. Distributors were directed to provide the recall information to their accounts.
Quantity
46 units
Distribution
Worldwide Distribution - US Nationwide in the states of UK, Canada, Holland
Lot, serial or model codes
(a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 Serial Numbers: 1000054866 1000054860 1000051368 1000054859 1000054855 1000054857  (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 Serial Numbers: 1000053319 1000053312 1000052886 1000053325 1000053323 1000052420 1000052259 1000052426 1000052425 1000052432 1000052891 1000052458 1000052429 1000052256 1000052888 1000053990 1000051418 1000051419 1000052457 1000052252 1000051411 1000055556 1000052460 1000054129 1000053979 1000049129 1000051412 1000052424 1000052455 1000054117 1000053982 1000053983 1000054122 1000054121  (c) PROCAIR PLUS 84" MATTRESS, MATERIAL NUMBER 222-6384 Serial Numbers: 1000055501 1000049132 1000055498 1000057147 1000055500 1000056219
510(k) numbers
not on file
PMA numbers
not on file
Product code
IKY
Firm FEI
3009402404
Firm city
Arlington
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28