Advanced Bionics, LLC: AB HiResolution Bionic Ear System, HiRes Ultra 3D CI, HiFocus SlimJ Electrode - Product Usage: The device is intended to restore a level of auditory sensation to patients with profound sensorineural deafness via electrical stimulation of the auditory nerve.

Recall number
Z-2032-2020
Recalling firm
Advanced Bionics, LLC
Product
AB HiResolution Bionic Ear System, HiRes Ultra 3D CI, HiFocus SlimJ Electrode - Product Usage: The device is intended to restore a level of auditory sensation to patients with profound sensorineural deafness via electrical stimulation of the auditory nerve.
Status
Terminated
Initiated
2020-02-17
Posted
not on file
Terminated
2023-01-26
Reason
hearing performance degradation due to body-fluid entering the device.
Root cause
Device Design
Action
On 02/17/2020, the firm sent an "Urgent Field Action Notification" via email to healthcare providers informing them that the firm has detected an increase in the number of devices being explanted or the potential of being explanted due to performance issues (impedance drop and reports of hearing performance degradation) resulting from fluid ingress at the electrode. The Healthcare Notification instructs Healthcare Providers to: 1. Provide registration information for all devices that have been implanted; 2. Return any unimplanted devices and to emphasize not to implant any of these units. The Healthcare Provider Notification also provided the Healthcare Provider with a Patient Management information that includes "Key Clinical Questions". For questions, contact the Recalling Firm at: In the United States: (877) 8290026 between the hours of 5:00 AM and 5:00 PM Pacific Time, Monday through Friday. Europe / Asia Pacific / Latin America / Middle East and Africa: Please contact your local Advanced Bionics office with any questions. On 02/17/2020, the firm sent an "Important Notification" to implant receipients via FedEx to inform implant users or parents/caregivers of the issue of prolonged hearing degradation due to physiological fluid entering into the electrode (not the hermetic seal of the device body) and causing interruption of stimulation that can negatively affect device performance. The notification also informed users that the firm, as a precaution, is removing from the market all the initial version of unimplanted device and states that the performance degradation does not affect the safety of the device but revision surgery may be necessary due to the performance decrease. The firm is informing patients/users: -They should continue to use your device as normal. If you experience a decrease in hearing performance over time, you should visit your Audiologist to have your hearing assessed to understand if there could be a potential issue with yo
Quantity
8159 units (CI-1601-04 = 3827 units and CI-1601-05 = 4332 units)
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Brazil, Bangladesh, Switzerland, China, Australia, India, South Korea, Thailand, Ecuador, Mexico, El Salvador, Guatemala, Malaysia, Myanmar, Taiwan, Chile, Costa Rica, Colombia, Panama, Japan, Argentina, Peru, Singapore, China, New Zealand, Vietnam, Dominican Republic Refer to uploaded consignee lists in Associated Documents: Exhibit 05_ConsigneeList_Center_17FEB2020 Exhibit 06_ConsigneeList_Professionals_17FEB2020 Exhibit 07_ConsigneeList_Patients_17RFEB2020
Lot, serial or model codes
Model Numbers: CI 1601 04 - HiRes Ultra 3D with HF Mid-Scala Electrode CI 1601 05 - HiRes Ultra 3D with HF SlimJ Electrode  Lot Numbers - 1000000 thru 1999999.
510(k) numbers
not on file
PMA numbers
["P960058"]
Product code
MCM
Firm FEI
1000220815
Firm city
Valencia
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register