FDA Medical Device Recalls (openFDA)
GE Medical Systems China Co., Ltd.: GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
- Recall number
- Z-2029-2025
- Recalling firm
- GE Medical Systems China Co., Ltd.
- Product
- GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
- Status
- Open, Classified
- Initiated
- 2025-05-16
- Posted
- 2025-06-26
- Terminated
- not on file
- Reason
- GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
- Root cause
- Software design (manufacturing process)
- Action
- GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 05/16/2025 via letter delivered using a traceable means. The notice explained the issue, potential risk, and requested the following: "Actions to be taken by Customer/User: You may continue to use your Versana Premier R3 and LOGIQ F R3 series ultrasound systems by following the actions below: 1) If you use the Whizz report feature, reboot the system after finishing each obstetric patient examination. OR 2) Instead of using the Whizz report feature, after completing an obstetric patient examination, print all obstetric measurement and calculation results from the worksheet directly. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.71082@gehealthcare.com." If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 34 units
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- UDI/DI 00195278722430, Lot/Serial Numbers: 6310028WX0, 6310008WX0, 6310041WX0, 6310017WX0, 6310031WX0, 6310044WX0, 6310024WX0, 6310038WX0, 6310040WX0, 6310019WX0, 6310018WX0, 6310023WX0, 6310033WX0, 6310012WX0, 6310026WX0, 6310020WX0, 6310032WX0, 6310014WX0, 6310021WX0, 6310025WX0, 6310010WX0, 6310027WX0, 6310037WX0, 6310009WX0, 6310007WX0, 6310035WX0, 6310016WX0, 6310039WX0, 6310036WX0, 6310043WX0, 6310011WX0, 6310015WX0, 6310013WX0, 6310029WX0.
- 510(k) numbers
- ["K242005"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 3003513129
- Firm city
- Wuxi
- Firm state
- XIN
- Firm country
- China
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28