DRE Medical Group Inc: Philips Respironics V60 Ventilator-Intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Model Number: R1053618 and R1076709. These are retrofitted device.

Recall number
Z-2025-2023
Recalling firm
DRE Medical Group Inc
Product
Philips Respironics V60 Ventilator-Intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Model Number: R1053618 and R1076709. These are retrofitted device.
Status
Open, Classified
Initiated
2023-05-15
Posted
2023-06-29
Terminated
not on file
Reason
Devices were imported with Japanese Language and NIST Oxygen Gas Fittings and retrofitted, may pose a serious health risk as the gas pathway of these devices may be contaminated.
Root cause
No Marketing Application
Action
Avante Health issued URGENT: Medical Device Recall Philips Respironics V60 Ventilators Letter on 5/12/23 via email. Letter states reason for recall, health risk and action to take: " Immediately discontinue use. " Contact Avante Health Solutions for an RMA and shipping label to return the device. These devices must be returned to our facility for destruction. " Comparable replacement devices will be available for immediate shipment or " Credits can be issued. " Please email quality.management@avantehs.com to acknowledge receipt of this urgent recall " Attachments of Acknowledgement and Receipt Forms . Contact Information: Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time. (800) 462-8195 or (502) 244-4444 Quality.management@avantehs.com
Quantity
22 units
Distribution
US: CA, FL
Lot, serial or model codes
USI-DI: 00884838009868 Serial Numbers: Serial Number 100043954 100050078 100052760 100052848 100054619 100054626 100055277 100055557 100058549 100058566 100058837 100058995 100060228 100065373 100069520 100070487 100071122 100072691 100078379 100080232 100080844 201003462
510(k) numbers
not on file
PMA numbers
not on file
Product code
MNT
Firm FEI
1000122786
Firm city
Louisville
Firm state
KY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register