GE OEC Medical Systems, Inc: OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical structures during open or percutaneous surgical procedures. The system is an aid to locate anatomical structures during open or percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgery technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.
- Recall number
- Z-2025-2012
- Recalling firm
- GE OEC Medical Systems, Inc
- Product
- OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical structures during open or percutaneous surgical procedures. The system is an aid to locate anatomical structures during open or percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgery technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.
- Status
- Terminated
- Initiated
- 2007-11-19
- Posted
- 2012-07-18
- Terminated
- 2013-01-28
- Reason
- Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.
- Root cause
- Device Design
- Action
- Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE Healthcare sent an"URGENT RECALL NOTICE" dated February 22, 2008 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact the firm at 800-874-7378 for questions concerning this recall.
- Quantity
- 726 units
- Distribution
- Worldwide Distribution.
- Lot, serial or model codes
- Part numbers: 00-887200-01; 00-887200-03; 00-887201-01; 00-887201-03; 00-887202-01; 00-887202-03; 00-887204-03; 00-887207-03; 00-887206-03. Along with all associated serial numbers.
- 510(k) numbers
- ["K041932"]
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 1720753
- Firm city
- Salt Lake City
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28