Siemens Medical Solutions USA, Inc: Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany 95478. The 550 TxT Treatment Table is to be used as an integral part of the Siemens linear accelerator systems to position the patient accurately according to the treatment plan for various radiation therapy treatment techniques.

Recall number
Z-2023-2011
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany 95478. The 550 TxT Treatment Table is to be used as an integral part of the Siemens linear accelerator systems to position the patient accurately according to the treatment plan for various radiation therapy treatment techniques.
Status
Terminated
Initiated
2010-03-31
Posted
2011-04-19
Terminated
2012-03-01
Reason
Potential for causing overheating of the power supply during treatment, and potential for unexpected table descent.
Root cause
Component design/selection
Action
Siemens issued a Customer Information letter dated March 21, 2010. There were two safety issues. There were two occurrences of smoke caused by an overheating of the power supply during patient treatment. Also, there had been a report of an unexpected lowering of the 550TxT table in patient rescue mode. With the released kit that accompanies this letter, Siemens has implemented a minor modification on the 550TxT to include an additional fuse within the power supply circuit to address the first issue described above. They also included an additional modification to address the lowering of the table by the integration of additional relays. As soon as the "move enable" button is released, the table brake will be enabled which is controlled by the implemented relays. No further action is required for the continued operations of the 550TxT table. Further questions regarding this recall please call (925) 246-8407
Quantity
326 units
Distribution
Worldwide Distribution -- Including USA, states of NJ, WI, NY, and MA.
Lot, serial or model codes
All units of Listing Number: R864406
510(k) numbers
["K050422"]
PMA numbers
not on file
Product code
JAI
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register