Randox Laboratories Ltd.: Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes. Catalog Number: IA2640
- Recall number
- Z-2021-2023
- Recalling firm
- Randox Laboratories Ltd.
- Product
- Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes. Catalog Number: IA2640
- Status
- Open, Classified
- Initiated
- 2023-05-04
- Posted
- 2023-06-28
- Terminated
- not on file
- Reason
- Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range
- Root cause
- Nonconforming Material/Component
- Action
- Randox issued via email to the Distributor Medical Device Correction Letter and forwarded to US and PR customers on 04 May, 2023. Letter states reason for recall, health risk and action to take: Discontinue use of and discard the current IFU and download the updated IFU from randox.com Discuss the contents of this notice with your Medical Director. Complete and return the response for 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization. If you have any questions or concerns please contact Randox Technical Services.
- Quantity
- 1 kit
- Distribution
- CA, IL, VA, WA, WI including PR
- Lot, serial or model codes
- GTIN:05055273203868 Batch/Lot Number: 2031EC Exp. Date: 28th Nov 23
- 510(k) numbers
- ["K140522"]
- PMA numbers
- not on file
- Product code
- JJY
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- North
- Firm country
- Ireland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28