Terumo Cardiovascular Systems Corporation: CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit

Recall number
Z-2018-2025
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
Status
Open, Classified
Initiated
2025-05-28
Posted
2025-06-24
Terminated
not on file
Reason
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
Root cause
Employee error
Action
On May 28, 2025, firm began notifying affected consignees via Urgent Medical Device Removal letters. A Terumo Service Representative will contact each customer to discuss removal and replacement of affected devices. The firm is not recommending that customers halt use of their CDI OneView BPM Probe while waiting for removal and replacement. In the event suspected parameter inaccuracy occurs, performance of an in vivo recalibration is recommended. Per the Operator's Manual, displayed values should be verified for accuracy from another source before initial treatment.
Quantity
28 US; 7 OUS
Distribution
US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.
Lot, serial or model codes
UDI-DI 00886799001889 Lot Number/Serial Number Range: B0050231-B0050240 B0050221-B0050230 B0050211-B0050219 B0050201-B0050209 B0050191-B0050200
510(k) numbers
["K234065"]
PMA numbers
not on file
Product code
DRY
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register