Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: Olympus EVIS EXERA III BRONCHOVIDEOSCOPE BF TYPE, Model Numbers BF-H190, BF-Q190, BF-XT190, BF-1TH190

Recall number
Z-2017-2023
Recalling firm
Olympus Corporation of the Americas
Product
Olympus EVIS EXERA III BRONCHOVIDEOSCOPE BF TYPE, Model Numbers BF-H190, BF-Q190, BF-XT190, BF-1TH190
Status
Open, Classified
Initiated
2023-06-08
Posted
2023-07-17
Terminated
not on file
Reason
There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Root cause
Under Investigation by firm
Action
An "URGENT MEDICAL DEVICE CORRECTIVE ACTION" notification letter dated 6/8/23 was sent to customers. Actions to be taken by the company: The labelling will be updated to include specificity about laser compatibility, improved instructions regarding patient preparation, and warnings about patient injury and death resulting from incompatible laser use. Actions to be taken by the end user: 1. Inspect your inventory for the referenced devices and identify any device with the model names. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum . The addendum provides compatible laser type. 3. Ensure all personnel are completely knowledgeable and thoroughly aware that Olympus laser compatible bronchoscopes are compatible only with Nd: YAG laser or 810 nm diode lasers. 4. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0427 5. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Olympus requests that you report complaints, including any injuries associated with laser procedures with Olympus bronchoscopes, to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Olympus regrets any inconvenience caused and fully appreciates your cooperation in this matter. Please do not hesitate to call (647) 999-3203 or email: at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Quantity
9,684 units
Distribution
US Nationwide Distribution.
Lot, serial or model codes
All serial numbers BF-H190 EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190 UDI: 04953170434754 BF-Q190 EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190 UDI: 04953170335198 BF-XT190 EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XT190UDI: 04953170402470 BF-1TH190 EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190 UDI: 04953170434778
510(k) numbers
["K121959","K183419"]
PMA numbers
not on file
Product code
EOQ
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register