FDA Medical Device Recalls (openFDA)
Medline Industries, LP: Medline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK, Medline SKU NEUROSPNG02; 2. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 0.5", XR, 10/PK, Medline SKU NEUROSPNG05; 3. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG06; 4. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1.5", XR, LF, 10/PK, Medline SKU NEUROSPNG07; 5. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG09; 6. Neuro Sponges, Eponges neuro, Neuroesponjas, size .75" X .75", XR, LF, 10/PK, Medline SKU NEUROSPNG11; 7. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG13; 8. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG14.
- Recall number
- Z-2008-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK, Medline SKU NEUROSPNG02; 2. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 0.5", XR, 10/PK, Medline SKU NEUROSPNG05; 3. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG06; 4. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 1.5", XR, LF, 10/PK, Medline SKU NEUROSPNG07; 5. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.5" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG09; 6. Neuro Sponges, Eponges neuro, Neuroesponjas, size .75" X .75", XR, LF, 10/PK, Medline SKU NEUROSPNG11; 7. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 1", XR, LF, 10/PK, Medline SKU NEUROSPNG13; 8. Neuro Sponges, Eponges neuro, Neuroesponjas, size 1" X 3", XR, LF, 10/PK, Medline SKU NEUROSPNG14.
- Status
- Open, Classified
- Initiated
- 2026-03-19
- Posted
- 2026-04-30
- Terminated
- not on file
- Reason
- During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
- Root cause
- Under Investigation by firm
- Action
- Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their customers and should include customer quantities on your response form. For consignees of Medline medical convenience kits containing the affected sponges the actions to be taken are as follows: Identify and quarantine all affected kits, complete response form using website link: https://recalls.medline.com. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors must notify their customers and should include customer quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 4225160 units
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- 1. NEUROSPNG02: UDI/DI each 10197344111572, UDI/DI case 20197344111579, ALL LOTS; 2. NEUROSPNG05: UDI/DI each 20080196286659, UDI/DI case 30080196286656, ALL LOTS; 3. NEUROSPNG06: UDI/DI each 20080196286505, UDI/DI case 30080196286502, ALL LOTS; 4. NEUROSPNG07: UDI/DI each 10080196286485, UDI/DI case 40080196286486, ALL LOTS; 5. NEUROSPNG09: UDI/DI each 20080196286536, UDI/DI case 30080196286533, ALL LOTS; 6. NEUROSPNG11: UDI/DI each 20080196286550, UDI/DI case 30080196286557, ALL LOTS; 7. NEUROSPNG13: UDI/DI each 10080196286577, UDI/DI case 40080196286578, ALL LOTS; 8. NEUROSPNG14: UDI/DI each 10080196286584, UDI/DI case 40080196286585, ALL LOTS.
- 510(k) numbers
- ["K140369"]
- PMA numbers
- not on file
- Product code
- HBA
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28