Ra Medical Systems, Inc.: Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE BY: 2019-09-24, MFR DATE: 2018-09-24, CE, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Revised Label: Product Label - Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 19-092400-041, USE BY: 2019-12-13, MFR DATE: 2019-10-14, CE 2797, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Package Label - DABRA 101 5 French, Ref: 1222-5000-01, Lot: LOT: 19-092400-041, USE BY: 2019-12-13, UDI's (01)00842362100031, (17)20191213(21)2019092400041

Recall number
Z-2006-2020
Recalling firm
Ra Medical Systems, Inc.
Product
Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE BY: 2019-09-24, MFR DATE: 2018-09-24, CE, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Revised Label: Product Label - Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 19-092400-041, USE BY: 2019-12-13, MFR DATE: 2019-10-14, CE 2797, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Package Label - DABRA 101 5 French, Ref: 1222-5000-01, Lot: LOT: 19-092400-041, USE BY: 2019-12-13, UDI's (01)00842362100031, (17)20191213(21)2019092400041
Status
Terminated
Initiated
2019-09-19
Posted
not on file
Terminated
2023-07-31
Reason
The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.
Root cause
Under Investigation by firm
Action
On 9/27/19, the firm sent a "Medical Device Recall" letter to its customers via first class mail and emailed to inform them they are issuing a voluntary recall of DABRA Catheters manufactured within the time period specified below. Copies of letters were provided to Ra Medical Systems' representatives, who will visit current customers. The reason for the recall is that data collected through customer feedback and field case reporting, corroborated by internal testing, suggest that the product shelf life is currently less than the labeled shelf life, which has resulted in a higher catheter calibration failure rate. The product impacted is the DABRA Catheter manufactured from 9/24/18 - 8/26/19. The firm stated that failure to calibrate does not pose a safety risk to patients or users. Ra Medical is not aware of any reports of injury or harm related to this issue, nor is there any safety concerns for patients that have been successfully treated using products defined within this recall. The firm is instructing customer to complete recall acknowledgement form that indicates the quantity of product they have. When the form is returned to Ra Medical, we will send the customer return packaging and a shipping label for them to return the recalled product to our facility. Ra Medical Systems representatives will continue to contact customers until all products are returned, or verified no longer at the customers' sites. Each returned product will have an RMA number assigned per our procedure. It will be tagged and stored in a quarantined area. The firm will contact the customers by phone, email or mail to ensure receipt and/or compliance. Should you have any questions or need further assistance please feel free to contact Ra Medical at (442) 325-5044
Quantity
5082 products
Distribution
Worldwide Distribution - US Nationwide - MI, TN, CA, FL, OR, PA, AZ, AL, VA, GA, TX, ID, AR, NC, IN, SC, CT, NY, KS, LA, NV, MS, NE, MD, IL, OK, MA, WA, MO, WI, KY. OUS - Netherlands, Pakistan, United Kingdom, Italy, France
Lot, serial or model codes
DABRA 101 Catheter.  Ref: 1222-5000-01 Serial Number(s): Manufacturing dates 2018-09-03 to 2019-08-26. Expiration dates: September 2019 to August 2020
510(k) numbers
["K170349"]
PMA numbers
not on file
Product code
PDU
Firm FEI
3004217728
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register