Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corporation: Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the hear-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours

Recall number
Z-2002-2016
Recalling firm
Datascope Corporation
Product
Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the hear-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours
Status
Terminated
Initiated
2016-05-09
Posted
not on file
Terminated
2017-03-28
Reason
Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.
Root cause
Under Investigation by firm
Action
Maquet sent via express mail and hand delivered a recall letter and response form dated May 9, 2016 to their sole customer.
Quantity
17 units
Distribution
US in the state of GA
Lot, serial or model codes
Lot/batch No: 3000003260, Part number 701050175
510(k) numbers
not on file
PMA numbers
not on file
Product code
DWE
Firm FEI
1000222374
Firm city
Fairfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register