FDA Medical Device Recalls (openFDA)
ICU Medical, Inc.: Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41414-23
- Recall number
- Z-2001-2020
- Recalling firm
- ICU Medical, Inc.
- Product
- Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 41414-23
- Status
- Terminated
- Initiated
- 2020-01-02
- Posted
- not on file
- Terminated
- 2022-09-14
- Reason
- Inability for the guidewire to pass through the needles included with the catheter kits.
- Root cause
- Process control
- Action
- The firm has contracted Stericycle, Inc., Indianapolis, IN for customer notification, response tracking and effectiveness check. On 02 January 2020, Stericycle, Inc. sent URGENT: MEDICAL DEVICE RECALL notification of the market action to the firm's direct consignees of record.
- Quantity
- 235
- Distribution
- US Consignees including in the states of AL, CA, FL, GA, IL, ME, MI, MN, MS, NC, NH, NM, NY, OH, TX, UT, VA and WA OUS - Canada.
- Lot, serial or model codes
- LOT Number/ UDI - 4156000/1 0840619 04647 4, 4174850/1 0840619 04647 4, 4192313/1 0840619 04647 4, 4330831/1 0840619 04647 4, 4365303/1 0840619 04647 4
- 510(k) numbers
- ["K091268"]
- PMA numbers
- not on file
- Product code
- DQE
- Firm FEI
- 2025816
- Firm city
- San Clemente
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28