Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: 00597001101 Femur CR PRECOAT FEM COMP SIZE AML, Rx, Sterile; 00597001102 Femur CR PRECOAT FEM COMP SIZE AMR, Rx, Sterile; 00597001201 Femur CR PRECOAT FEM COMP SIZE BML, Rx, Sterile; 00597001202 Femur CR PRECOAT FEM COMP SIZE BMR, Rx, Sterile. Used in total knee arthroplasty and indicated for patients with severe pain.

Recall number
Z-1998-2013
Recalling firm
Zimmer, Inc.
Product
00597001101 Femur CR PRECOAT FEM COMP SIZE AML, Rx, Sterile; 00597001102 Femur CR PRECOAT FEM COMP SIZE AMR, Rx, Sterile; 00597001201 Femur CR PRECOAT FEM COMP SIZE BML, Rx, Sterile; 00597001202 Femur CR PRECOAT FEM COMP SIZE BMR, Rx, Sterile. Used in total knee arthroplasty and indicated for patients with severe pain.
Status
Terminated
Initiated
2013-06-07
Posted
2013-08-22
Terminated
2014-12-15
Reason
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Root cause
Labeling design
Action
Beginning June 7, 2013 all distributors will be sent Urgent Device Recall notices via electronic mail, with distributors, hospital risk managers with affected inventory also receiving the notice via courier. The Notice outlined the products recalled; the reason for the recall, possible adverse events related to the recall and directed consignees to locate, quarantine and count all affected product recording the information on the Inventory Return Form acknowledging that recall directions were followed. Further questions should be directed to 1-877-946-2761. Affected product is to be returned to Zimmer Product Service Department , 1777 West Center St. Warsaw, IN 46580 or internation return at Zimmer International Logistics GmbH, Max - Immelmann Allee 12 79427 Eschbach Germany. " Communications outside of the United States will occur approximately two weeks after the United States communications.
Quantity
192,355 all devices
Distribution
Worldwide Distribution-USA (nationwide) and the country of Japan.
Lot, serial or model codes
REF 00597001101, all codes
510(k) numbers
not on file
PMA numbers
not on file
Product code
JWH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register