Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Access Systems Inc.: 5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF CK000340B BARD MYPICC KIT 5F French Size Dual-Lumen 70cm Guidewire

Recall number
Z-1994-2021
Recalling firm
Bard Access Systems Inc.
Product
5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF CK000340B BARD MYPICC KIT 5F French Size Dual-Lumen 70cm Guidewire
Status
Terminated
Initiated
2021-05-19
Posted
not on file
Terminated
2023-03-31
Reason
Catheter convenience kits contain the incorrect guidewire. Kits were labeled with 70cm guidewires, but contained a 50cm guidewire.
Root cause
Process control
Action
On 05/19/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE CORRECTION" Notification via FedEx 2-day to customers informing them that the Recalling Firm has confirmed through customer complaint that a number of catheter convenience kits labeled as having 70cm guidewires may actually contain 50cm guidewires. The Recalling Firm is providing customers with a quick way of identifying if their kit contains the correct or incorrect size guidewire by determining the number of loops of the wire. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and lot numbers listed. Determine if the product has the correct product based on the number of loops of the guidewire while the product is still in the packaging material. Product may be discarded when clinically necessary and replacement requested. 2. Share this recall notification with all users within or outside their facility network to which they may have further distributed the affected products to ensure these additional users are also aware of this advisory notice. 3. Complete the attached Customer Response Form and return to the Recalling Firm contact noted on the form whether or not they have any of the impacted material so that Recalling Firm may acknowledge receipt of this notification per FDA requirements and subsequently process the product replacement, if necessary. For further assistance or for questions - contact: North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted MonFri 8:00am and 5:00pm CT
Quantity
30 kits
Distribution
U.S. Nationwide distribution in the state of RI.
Lot, serial or model codes
Catalog (REF) Number: CK000340B UDI Code: (01)00801741126697 Lot Number: 20BBK658
510(k) numbers
not on file
PMA numbers
not on file
Product code
LJS
Firm FEI
3006260740
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register