FDA Medical Device Recalls (openFDA)

Getinge USA Inc: Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623. Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121C, 132C, 134C and 135C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.

Recall number
Z-1994-2013
Recalling firm
Getinge USA Inc
Product
Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623. Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121C, 132C, 134C and 135C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.
Status
Terminated
Initiated
2013-08-02
Posted
2013-08-21
Terminated
2017-02-23
Reason
AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization parameters.
Root cause
Other
Action
GETINGE sent an DEVICE CORRECTION NOTICE letter dated July 31, 2013, and response forms were sent to customers on August 2, 2013, via Certified Mail Return Receipt Requested. The letters instruct customers to immediately discontinue use, quarantine, and destroy the affected lot. In addition, they instruct customers to: Check the appropriate box and record the quantity of affected unused product on the enclosed Correction Response Form OR If if they do not have any lots of affected unused inventory check the box indicating such. Destroy the unused product in accordance with your institution's policy and instructions for use. Record the method of destruction on the Correction Response Form. Be sure to include the Facility Name, Address, Contact Name and Phone Number. Sign, date and e-mail the form back to SR430responseform@getinge.com within 10 days or sooner of receiving this notice. Credit will be issued upon receipt of the Correction Response Form. Replacement product of unaffected lots is currently available. The warehouse shipping the affected lot is no longer being used (due to business decision, not as result of this recall). The current warehouse is maintaining the product in a controlled temperature segregated room away from any potential sterilants. If you have any questions regarding this notification please call 1-800-475-1400 ext. 5026
Quantity
Catalog # 61301606636: 78 Packs (3900 BIs); Catalog # 61301606637: 43 Packs (4300 BIs)
Distribution
Worldwide Distribution - USA and the country of Canada.
Lot, serial or model codes
LOT SR-430 (Expiration 02/06/2014)
510(k) numbers
["K093794"]
PMA numbers
not on file
Product code
FRC
Firm FEI
3004147784
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register