Fresenius Medical Care Holdings, Inc.: 5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.

Recall number
Z-1992-2025
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.
Status
Open, Classified
Initiated
2025-04-29
Posted
2025-06-18
Terminated
not on file
Reason
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
Root cause
Process control
Action
Fresenius Medical Care notified its single consignee on 04/29/2025 via telephone call. The consignee was made aware of the issue and affected units and instructed to examine their inventory for those affected units. If they had any affected units, they were instructed to remove them from their inventory immediately and provide contact information to arrange for their return. On 05/14/2025, Fresenius Medical Care expanded the recall to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. On 06/05/2025, Fresenius Medical Care expanded the recall again to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. A recall notification letter was sent to the consignee on about 06/23/2025. The letter provided the same information as was verbally communicated in notifications.
Quantity
Model Number, Cases (Eaches): 03-5100-7, 76 (1,824); 03-5100-7C, 5 (120); 03-5150-2, 8 (192)
Distribution
US Nationwide distribution in the state of Massachusetts.
Lot, serial or model codes
a. Model Number: 03-5100-7; UDI-DI: 00840861102365 (Bag), 10840861102362 (Case); Lot Numbers: 24LR01214, 24SR01010, 25ER01800. b. Model Number: 03-5100-7C; UDI-DI: 0084086110237 (Bag), 10840861102379 (Case); Lot Number: 24LR01215, 24SR01011, 25ER01804. c. Model Number: 03-5150-2; UDI-DI: 00840861102389 (Bag), 10840861102386 (Case); Lot Number: 24LR01216, 24SR01012, 25ER01805.
510(k) numbers
["K231534"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register