Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, Lp: Turon Revision Glenoid Poly Liner. The Metal Backed Revision Glenoid is a two piece metal backed polyethylene glenoid component for use in total shoulder arthroplasty. It includes a metal baseplate and polyethylene liner. DJO Surgical shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the indications for use.

Recall number
Z-1992-2009
Recalling firm
Encore Medical, Lp
Product
Turon Revision Glenoid Poly Liner. The Metal Backed Revision Glenoid is a two piece metal backed polyethylene glenoid component for use in total shoulder arthroplasty. It includes a metal baseplate and polyethylene liner. DJO Surgical shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the indications for use.
Status
Terminated
Initiated
2009-07-20
Posted
2009-09-04
Terminated
2010-01-07
Reason
Product devices used for testing are not identical to what was manufactured. Under certain circumstances snap feature could allow poly liner to move on baseplate reducing tightness of fit.
Root cause
Release of Material/Component prior to receiving test results
Action
Firm notified consignees by an Urgent: Medical Device Recall letter, dated July 20, 2009. The letter stated the reason for the recall, asked customers to quarantine, and return the affected product. The enclosed response form needed to be completed and returned as soon as possible. Questions were to be directed to David Harris, Manager of Customer Service at 512-834-6330.
Quantity
60 units
Distribution
Nationwide Distribution -- Including states of TX, FL, and GA.
Lot, serial or model codes
Lot numbers: 53987943, 53987944, 53987945, 53987946, and 53987947.
510(k) numbers
["K081448"]
PMA numbers
not on file
Product code
MBF
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register