ICU Medical, Inc.: ICU Medical Plum Solos, Item number 400011001; infusion pump

Recall number
Z-1991-2025
Recalling firm
ICU Medical, Inc.
Product
ICU Medical Plum Solos, Item number 400011001; infusion pump
Status
Open, Classified
Initiated
2025-04-22
Posted
2025-06-18
Terminated
not on file
Reason
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.
Root cause
Under Investigation by firm
Action
ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 05/02/2025 via hand delivery. The notice explained the issue, potential risk, and requested the following: Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification. 2. If Maximum Dose Limit is not used for a medication, the issue will not occur. If Maximum Dose Limit is used for a medication, users should take the following precautions. If you need to adjust the weight or BSA, first clear the program and then reprogram. For further inquiries: To report adverse events or product complaints: Global Complaint Management - 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or ProductComplaintsPP@icumed.com. For any questions regarding this action: Device Correction Inquiries https://icumed.custhelp.com/app/market-action. Additional technical information or assistance: Technical Support 1-(800)-241-4002, option 3 or tsc.support@icumed.com.
Quantity
214 units
Distribution
Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.
Lot, serial or model codes
UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 60000114, 60000120, 60000130, 60000133, 60000134, 60000136, 60000139, 60000141, 60000143, 60000147, 60000150, 60000154, 60000159, 60000160, 60000162, 60000165, 60000167, 60000170, 60000171, 60000172, 60000173, 60000174, 60000176, 60000177, 60000178, 60000179, 60000180, 60000183, 60000187, 60000189, 60000190, 60000191, 60000192, 60000198, 60000199, 60000201, 60000203, 60000204, 60000205, 60000206, 60000207, 60000208, 60000209, 60000210, 60000212, 60000213, 60000214, 60000215, 60000216, 60000217, 60000223, 60000224, 60000225, 60000226, 60000227, 60000228, 60000229, 60000230, 60000234, 60000236, 60000237, 60000238, 60000239, 60000240, 60000241, 60000242, 60000244, 60000245, 60000246, 60000247, 60000249, 60000251, 60000252, 60000255, 60000257, 60000259, 60000263, 60000264, 60000265, 60000266, 60000268, 60000269, 60000271, 60000273, 60000274, 60000275, 60000276, 60000277, 60000279, 60000280, 60000281, 60000282, 60000283, 60000284, 60000285, 60000286, 60000288, 60000291, 60000292, 60000295, 60000296, 60000297, 60000298, 60000300, 60000155, 60000193, 60000197, 60000211, 60000243, 60000270, 60000272, 60000123, 60000168, 60000175, 60000195, 60000221, 60000256, 60000294.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FRN
Firm FEI
3013319212
Firm city
Lake Forest
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register