Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien Llc: Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Recall number
Z-1990-2020
Recalling firm
Covidien Llc
Product
Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Status
Terminated
Initiated
2020-04-30
Posted
not on file
Terminated
2023-12-14
Reason
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Root cause
Nonconforming Material/Component
Action
Cardinal Health issued letter Urgent Medical Device Recall EVENT-2020-02846 on 4/30/20 via overnight mail. Letter states reason for recall, health risk, and action to take: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers 2. SEGREGATE and QUARANTINE all on-hand product that is confirmed to be labeled per Exhibit A. 3. PLEASE RETURN the enclosed acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 5. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: Hospital800-964-5227 Federal Government800-444-1166 Distributor800-635-6021 All other Customers888-444-5440 Questions contact: Cardinal Health 800-292-9332
Quantity
23587 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Lot Numbers: 1826714264, 1908533364, 1909302964, 1909303064,  1910654864, 1911300664, 1916240464, 1921106564, 1922533864, 2004908564, 2005617264, 2006339564
510(k) numbers
["K933400"]
PMA numbers
not on file
Product code
KOD
Firm FEI
1282497
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register