Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.

Recall number
Z-1985-2014
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.
Status
Terminated
Initiated
2014-05-09
Posted
2014-07-02
Terminated
2014-12-15
Reason
There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.
Root cause
Component design/selection
Action
Siemens sent a Safety Advisory Notice letter dated May 9, 2014, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that a Siemens service technician would be visiting their site to identify and service affected devices.
Quantity
10
Distribution
US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.
Lot, serial or model codes
model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022
510(k) numbers
["K033238"]
PMA numbers
not on file
Product code
IZL
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register