Skip to the content

FDA Medical Device Recalls (openFDA)

Medicel Ag: Product Name: nanoPOINT Injector System, item number 37-0000-08. Intended to fold and insert a STAAR Surgical Collamer nanoFLEX or Afinity IOLs, through surgical procedure in a human eye.

Recall number
Z-1984-2013
Recalling firm
Medicel Ag
Product
Product Name: nanoPOINT Injector System, item number 37-0000-08. Intended to fold and insert a STAAR Surgical Collamer nanoFLEX or Afinity IOLs, through surgical procedure in a human eye.
Status
Terminated
Initiated
2013-08-05
Posted
2013-08-20
Terminated
2014-03-11
Reason
The firm initiated this recall of one lot (121402) of nanoPOINT Injectors after they noticed that Tyvek lids might have been damaged and, therefore, the stability of the sterile barrier might be negatively affected.
Root cause
Use error
Action
Medicel AG sent recall notification to its one direct consignee via email on 08/05/2013. The recall notification letter was titled "URGENT-Medical Device Correction Notice for nanoPOINT Injectors LOT 121402". The recall notification letter informed the customer of the product device description and lot number recalled, reason for recall, instructions to "not use this affected lot...and return them", contact information and an area on the bottom of the letter to acknowledge receipt of the correction notice. The firm's direct consignee who further distributed this product to the end users will send out notifications.
Quantity
68 boxes or 680 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Item number 37-0000-08  Lot number of the recalled products: LOT121402
510(k) numbers
["K070669"]
PMA numbers
not on file
Product code
HQL
Firm FEI
3006151006
Firm city
Wolfhalden
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register