Skip to the content

FDA Medical Device Recalls (openFDA)

Gyrus Medical, Inc: GYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting Forceps 5mm/45 cm, Sterile EO. GYRUS ACMI, REF 3740PK, or REF 94201PK, PKS LYONS, Dissecting Forceps 5mm/45cm, Sterile EO. The Dissectors are sterile, single use electrosurgical devices that are indicated for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general surgical procedures.

Recall number
Z-1983-2013
Recalling firm
Gyrus Medical, Inc
Product
GYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting Forceps 5mm/45 cm, Sterile EO. GYRUS ACMI, REF 3740PK, or REF 94201PK, PKS LYONS, Dissecting Forceps 5mm/45cm, Sterile EO. The Dissectors are sterile, single use electrosurgical devices that are indicated for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general surgical procedures.
Status
Terminated
Initiated
2013-07-24
Posted
2013-08-19
Terminated
2014-06-20
Reason
On April 29, 2013 Olympus Surgical Technologies America R&D received results from an accelerated aging study conducted as part of a new market submission. The study found that one of 60 Dissector units that underwent 3 year accelerated aging failed the dye penetration test indicating that the seal on sterile packaging was compromised. The dissector products are labeled with a 3 year shelf life.
Root cause
Packaging process control
Action
Olympus sent a "Medical Device Recall" letter dated July 24, 2013, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer.. Customers were advised to immediately cease any further use of the product and remove it from inventory and quarantine it until is shipped back to the company. To return the product, call 1-888-524-7266 Olympus customer service representative. Customers were requested to complete and return by fax the enclosed questionnaire to 484-896-7128. If the product was further distributed then they need to identify their customers and notify them at once of the product recall.
Quantity
1140
Distribution
Worldwide Distribution - USA (nationwide) including the states of AL, AZ, CA, FL, GA, IL, IN, KS,KY, MD, MA, MI, MN, MS, MO, NY, NC, OH, PA, SD, TN, TX, VA and WI., and the countries of BRAZIL, CHINA, AUSTRALIA, CHINA, GERMANY and INDIA.
Lot, serial or model codes
REF 3740  150080AC 258624MC G166568CC 381809KD 408020AE JF480565  NY492562  REF 3740PK 128971JB 226730IC 226742IC 226743IC 431894CE 431895CE G128677JB  REF 94201PK  102290EB 147428AC 167021CC 198596GC 204439HC 204451HC 249022KC 278142BD 320659DD 344712FD 363953ID 373761JD 389095MD 418231BE G386412KD G461037EE JF493195
510(k) numbers
["K031080"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3011050570
Firm city
Osseo
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register