Skip to the content

FDA Medical Device Recalls (openFDA)

Endogastric Solutions Inc: EndoGastric Solutions EsophyX2 Device with Serosa Fastener and Accessories. Models R2001 and R2002. Indicated for the treatment of symptomatic chronic gastroesophageal reflux disease.

Recall number
Z-1982-2013
Recalling firm
Endogastric Solutions Inc
Product
EndoGastric Solutions EsophyX2 Device with Serosa Fastener and Accessories. Models R2001 and R2002. Indicated for the treatment of symptomatic chronic gastroesophageal reflux disease.
Status
Terminated
Initiated
2013-06-06
Posted
2013-08-17
Terminated
2014-04-09
Reason
Endogastric Solutions, Inc. has received a limited number of reports relating to the loss of tissue mold control when operating the R2001 or R2002 EsophyX2 device. In one case, surgical intervention was necessary to remove the device.
Root cause
Nonconforming Material/Component
Action
Endogastric sent the Safety Alert: EndoGastric Solutions (EGS), EsophyX2 Device letter, dated June 5, 2013, to their consignees. Endogastric sent the second letter URGENT: MEDICAL DEVICE RECALL EsophyX2, dated July 26, 2013. This letter advised customers that the firm is voluntarily recalling EsophyX2 Devices with SerosaFuse Implantable Fastener and Accessories (Models R2001 and R2002) manufactured before February 2012. Customers who have UN-EXPIRED devices in their inventory are advised to discontinue use and fill out the Medical Device Recall Return Response form and return it to Endogastric. The Customer Service will contact customers with instructions on how to return the product to the company. Customers who may have EXPIRED devices are advised to with their central supply departments to ensure all identified products have been removed from inventory and destroyed. They should fill out the Medical Device Recall Return Response form with the lot numbers and quantity destroyed and return the completed form to the firm. Customers can call the Customer Service and Support at 425-307-9269, Monday through Friday, 8:00AM to 5:00 PM, Pacific Time for questions. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Quantity
5192 units in the US and 41 units outside the US
Distribution
Distributed nationwide and Italy.
Lot, serial or model codes
UN-EXPIRED devices with Expiration dates (8/31/2013, 11/30/2013, 12/31/2013, and 1/31/2014) and following LOT numbers:  401132, 401147, 401152, 401156, 401159, 401167, 401207, 401214, 401219, 401220, 401225, 401227, 401230, 401233, 401236, 401246, 401248, 401251, 401257, 401258, 401259, 401267, 401269, 401271, 401274, and 401276.   EXPIRED devices with Expired dates (1/31/2013, 2/28/2013, 3/31/2013 4/30/2013, 5/31/2013, 6/30/2013, and 7/31/2013) and following LOT numbers:  400923, 400927, 400934, 400937, 400940, 400943, 400949, 400961, 400963, 400973, 400976, 400979, 400980, 400981, 400984, 400986, 400992, 400994, 400997, 400998, 401004, 401005, 401009, 401012, 401016, 401017, 401024, 401026, 401029, 401056, 401058, 401059, 401060, 401061, 401062, 401063, 401065, 401072, 401074, 401076, 401079, 401084, 401085, 401092, 401094, 401107, 401114, 401115, 401121, 401129, and 401130.
510(k) numbers
["K092400"]
PMA numbers
not on file
Product code
ODE
Firm FEI
3005473391
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register