Skip to the content

FDA Medical Device Recalls (openFDA)

American Contract Systems Inc: Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ

Recall number
Z-1981-2026
Recalling firm
American Contract Systems Inc
Product
Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ
Status
Open, Classified
Initiated
2026-03-19
Posted
2026-05-06
Terminated
not on file
Reason
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
Root cause
Nonconforming Material/Component
Action
Firm notified distributors with an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" dated March 19, 2026, via email. Distributors were instructed to notify end-users. An updated letter with additional kits was sent to the customer on 3/20/26. Customers should quarantine all affected product. Customers may add warning labels to affected ACS kits to ensure all users are aware of prescribed actions that were identified in the Medline recall notice. The warning labels state, "Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure." Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. If you have any questions, contact the Regulatory Affairs Manager, by email: gm-omra_recalls@owens-minor.com or at 419-455-2090.
Quantity
7,311 kits
Distribution
US distribution to MN. No OUS distribution.
Lot, serial or model codes
ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
3008492337
Firm city
Tiffin
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register