FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-2 (straight handle) Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: PathFinder NXT is a minimally invasive pedicle screw system. The PathFinder NXT System consists of various screws and rods intended to provide temporary stabilization following surgery to fuse the spine. The intended use of the Fixed Angled Percutaneous Rod Inserter is to insert a rod percutaneously through the extender sieeves to position into the implanted screw heads for securing with closure tops forming a construct that provides temporary stabilization following MIS or mini-open surgery to fuse the spine.
- Recall number
- Z-1981-2012
- Recalling firm
- Zimmer, Inc.
- Product
- Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-2 (straight handle) Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: PathFinder NXT is a minimally invasive pedicle screw system. The PathFinder NXT System consists of various screws and rods intended to provide temporary stabilization following surgery to fuse the spine. The intended use of the Fixed Angled Percutaneous Rod Inserter is to insert a rod percutaneously through the extender sieeves to position into the implanted screw heads for securing with closure tops forming a construct that provides temporary stabilization following MIS or mini-open surgery to fuse the spine.
- Status
- Terminated
- Initiated
- 2012-06-05
- Posted
- 2012-07-10
- Terminated
- 2014-09-03
- Reason
- Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
- Root cause
- Device Design
- Action
- Zimmer sent an Urgent Medical Device Correction and Removal notification dated June 5, 2012 to distributors and all affected accounts of Zimmer Spine. The notice identified the affected products, problem and actions to be taken. Distributors were asked to hand deliver Recall Notices to Surgeons and return the completed Surgeon Contact Certification Form within 10 working days to ronald.musselman@zimmer.com or FAX to 1-512-258-0995. Notices to Surgeons and User facilities included revised directions/precautions for the Instructions for Use and attached Surgical Techique. Devices will be removed once newly design parts become available. For any related questions or assistance about this please contact Zimmer Spine Customer Service at 866-774-6368.
- Quantity
- 249 ( angle handle and straight) in total
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
- Lot, serial or model codes
- Part numbers 3573-2 (straight handle) all lots.
- 510(k) numbers
- ["K100845"]
- PMA numbers
- not on file
- Product code
- NKB
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as