Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

Recall number
Z-1980-2026
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Status
Open, Classified
Initiated
2026-03-18
Posted
2026-04-28
Terminated
not on file
Reason
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Root cause
Under Investigation by firm
Action
Siemens Healthineers notified consignees on about 03/18/2026 via letter. Consignees were notified of the issue, the hazard involved and instructed to immediately discontinue use of and discard the affected kit lots, each facility should review the recall notification with their Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, and review inventory to determine the laboratory's replacement needs and communicate that with Siemens Healthineers. Consignees were also instructed to complete and return the Field Correction Effectiveness Check Form provided and retain the letter with laboratory records and forward it to anyone how may have received or used the affected product.
Quantity
18,080 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.
Lot, serial or model codes
Material Number: 10872079. Model Number: DF33B. UDI: (01)00630414595009(10)GA6307(17)20261103, (01)00630414595009(10)BA7005(17)20270105. Lot Numbers: GA6307, BA7005.
510(k) numbers
["K132638"]
PMA numbers
not on file
Product code
CGX
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register