Skip to the content

FDA Medical Device Recalls (openFDA)

Ohio Medical Corporation: GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Recall number
Z-1980-2025
Recalling firm
Ohio Medical Corporation
Product
GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S
Status
Open, Classified
Initiated
2025-05-08
Posted
2025-06-13
Terminated
not on file
Reason
Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.
Root cause
Device Design
Action
Ohio Medical issued a Field Safety Notice to its consignees on 05/08/2025 followed by telephone calls on 05/13/2025. The notice explained the problem with the device, potential risk, and requested the devices be returned for correction. The notice further stated: "The Field Safety Notice is communicated to all impacted customers and is communicated to competent authorities in the affected markets. Receivers should pass this FSN on to all persons with a need to be advised from within your organization or to any organization where the potentially affected products have been supplied. Please consider end users and others within the supply chain in the circulation of this notice."
Quantity
36 units
Distribution
US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.
Lot, serial or model codes
Serial Numbers: ZL316367, ZL316446, ZL316462, ZL316475, ZL316481, ZL316484, ZL316490, ZL316493, ZL316503, ZL317057, ZL317087, ZL317088, ZL317090, ZL322386, ZL322387, ZL322390, ZL322392, ZL322409, ZL322412, ZL322439, ZL322441, ZL322442, ZL322443, ZL322445, ZL322447, ZL322451, ZL322455, ZL322456, ZL322462, ZL322463, ZL322466, ZL322467, ZL322470, ZL322472, ZL322473, ZL322837.
510(k) numbers
["K162433"]
PMA numbers
not on file
Product code
CAW
Firm FEI
1419185
Firm city
Gurnee
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register