Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);

Recall number
Z-1980-2012
Recalling firm
Zimmer, Inc.
Product
Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);
Status
Terminated
Initiated
2012-06-05
Posted
2012-07-10
Terminated
2014-09-03
Reason
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
Root cause
Device Design
Action
Zimmer sent an Urgent Medical Device Correction and Removal notification dated June 5, 2012 to distributors and all affected accounts of Zimmer Spine. The notice identified the affected products, problem and actions to be taken. Distributors were asked to hand deliver Recall Notices to Surgeons and return the completed Surgeon Contact Certification Form within 10 working days to ronald.musselman@zimmer.com or FAX to 1-512-258-0995. Notices to Surgeons and User facilities included revised directions/precautions for the Instructions for Use and attached Surgical Techique. Devices will be removed once newly design parts become available. For any related questions or assistance about this please contact Zimmer Spine Customer Service at 866-774-6368.
Quantity
249 ( angle handle and straight) in total
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
Lot, serial or model codes
Part numbers 3573-1 (angle handle) all lots including 64PY, 67YA, 69FF, 71GB, 72JL, 73GY
510(k) numbers
["K100845"]
PMA numbers
not on file
Product code
NKB
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28