Smiths Medical ASD, Inc.: Smiths Medical HOTLINE 3 Blood and Fluid Warmer, Reorder Numbers: HL-390 Product Usage: The Micro-switch is used on the HOTLINE 3 Blood and Fluid Warmer. This product, for use in conjunction with HOTLINE 3 L-370 Fluid Warming Set are intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures. These products are intended for use by appropriately trained healthcare professionals in clinical environments

Recall number
Z-1979-2012
Recalling firm
Smiths Medical ASD, Inc.
Product
Smiths Medical HOTLINE 3 Blood and Fluid Warmer, Reorder Numbers: HL-390 Product Usage: The Micro-switch is used on the HOTLINE 3 Blood and Fluid Warmer. This product, for use in conjunction with HOTLINE 3 L-370 Fluid Warming Set are intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures. These products are intended for use by appropriately trained healthcare professionals in clinical environments
Status
Terminated
Initiated
2012-06-27
Posted
2012-07-09
Terminated
2014-02-20
Reason
HOTLINE 3 Blood and Fluid Warmer( Reorder No. HL-390 ) disposable is not correctly installed , the microswitch may become damaged and unusable
Root cause
Device Design
Action
Smiths Medical sent an Urgent Medical Device Correction -Field Safety Notice letter dated June 21, 2012 via U.S. Mail to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter included an Urgent Medical Device Field Safety Notice Acknowledgement Form ("Confirmation Form") that customers were instructed to complete and return to Smiths Medical for carrying out the correction and affectivity tracking purposes. The letter stated that a Smiths Medical Representative will visit the affected sites to correct the units with a shelf upgrade kit, which includes instructions and all the parts necessary to make the correction to their affected devices. Customers were provided with a form to record the serial numbers of the devices corrected so that this can be returned to Smiths Medical. For questions regarding this information call Smiths Medical Quality Department at 1-781-763-9335. .
Quantity
2247 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
All serial numbers
510(k) numbers
["K911383"]
PMA numbers
not on file
Product code
LGZ
Firm FEI
1221261
Firm city
Rockland
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register