NeuMoDx Molecular Inc: NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100
- Recall number
- Z-1978-2021
- Recalling firm
- NeuMoDx Molecular Inc
- Product
- NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100
- Status
- Terminated
- Initiated
- 2021-05-04
- Posted
- not on file
- Terminated
- 2022-06-21
- Reason
- There is a potential for false positive results when certain lots of cartridges are used in conjunction with specific assays.
- Root cause
- Material/Component Contamination
- Action
- URGENT MEDICAL DEVICE CORRECTION notification letters dated 5/4/21 were sent to customers. Actions Required by Customers 1. IMMEDIATELY discontinue use of the product and discard remaining inventory in accordance with your national and local safety and environmental regulations. Please contact QIAGEN Technical Service for a free-of-charge replacement of scrapped inventory. 2. Share this Medical Device Correction Notice with all users of the NeuMoDxTM Cartridges in your facility to ensure they are aware of this notice. 3. If you have already used NeuMoDxTM Cartridges from any of these lots in combination with the NeuMoDxTM SARS-CoV-2 Assay or the NeuMoDx Flu-A-B/RSV/SARS-CoV-2 Assay, please review all SARS-CoV-2 positive results to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. Note: results for targets other than SARS-CoV-2 are not impacted by this medical device correction. 4. Complete the attached response form and return to Quality.Communications@qiagen.com to confirm receipt of this notification. We sincerely apologize for any inconvenience this may have caused and thank you in advance for your patience and cooperation. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com
- Quantity
- 620 cases (29,760 cartridges total)
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- GTIN: 10814278020274; Lot Numbers: 106629 106630 106631 106632
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- QJR
- Firm FEI
- 3010729479
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28