FDA Medical Device Recalls (openFDA)

NeuMoDx Molecular Inc: NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100

Recall number
Z-1978-2021
Recalling firm
NeuMoDx Molecular Inc
Product
NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100
Status
Terminated
Initiated
2021-05-04
Posted
not on file
Terminated
2022-06-21
Reason
There is a potential for false positive results when certain lots of cartridges are used in conjunction with specific assays.
Root cause
Material/Component Contamination
Action
URGENT MEDICAL DEVICE CORRECTION notification letters dated 5/4/21 were sent to customers. Actions Required by Customers 1. IMMEDIATELY discontinue use of the product and discard remaining inventory in accordance with your national and local safety and environmental regulations. Please contact QIAGEN Technical Service for a free-of-charge replacement of scrapped inventory. 2. Share this Medical Device Correction Notice with all users of the NeuMoDxTM Cartridges in your facility to ensure they are aware of this notice. 3. If you have already used NeuMoDxTM Cartridges from any of these lots in combination with the NeuMoDxTM SARS-CoV-2 Assay or the NeuMoDx Flu-A-B/RSV/SARS-CoV-2 Assay, please review all SARS-CoV-2 positive results to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. Note: results for targets other than SARS-CoV-2 are not impacted by this medical device correction. 4. Complete the attached response form and return to Quality.Communications@qiagen.com to confirm receipt of this notification. We sincerely apologize for any inconvenience this may have caused and thank you in advance for your patience and cooperation. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com
Quantity
620 cases (29,760 cartridges total)
Distribution
Worldwide distribution.
Lot, serial or model codes
GTIN: 10814278020274;  Lot Numbers: 106629 106630 106631 106632
510(k) numbers
not on file
PMA numbers
not on file
Product code
QJR
Firm FEI
3010729479
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register