FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: Integra LED Battery Charger, Single Bay. Product Number: 90523 The LED Headlight System Battery Charger (90523) is an accessory to the LED Headlight System: 90520US - LED Headlight w/Battery & AC/DC Power Supply - US 90520EU - LED Headlight w/Battery & AC/DC Power Supply - EU 90520UK - LED Headlight w/Battery & AC/DC Power Supply - UK 90520AU - LED Headlight w/Battery & AC/DC Power Supply - AU Product Usage: The Integra LED Headlight System Is designed to provide illumination to aid visualization during minor surgical, diagnostic, or therapeutic procedures
- Recall number
- Z-1976-2014
- Recalling firm
- Integra LifeSciences Corp.
- Product
- Integra LED Battery Charger, Single Bay. Product Number: 90523 The LED Headlight System Battery Charger (90523) is an accessory to the LED Headlight System: 90520US - LED Headlight w/Battery & AC/DC Power Supply - US 90520EU - LED Headlight w/Battery & AC/DC Power Supply - EU 90520UK - LED Headlight w/Battery & AC/DC Power Supply - UK 90520AU - LED Headlight w/Battery & AC/DC Power Supply - AU Product Usage: The Integra LED Headlight System Is designed to provide illumination to aid visualization during minor surgical, diagnostic, or therapeutic procedures
- Status
- Terminated
- Initiated
- 2014-06-11
- Posted
- 2014-07-02
- Terminated
- 2014-12-19
- Reason
- LED battery chargers may prematurely fail and will not charge the LED battery as intended
- Root cause
- Manufacturing material removal
- Action
- Integra initiated a voluntary recall on June 11, 2014 to consignees that have been shipped affected product lots of Integra LED Battery Chargers, Catalogue No. 90523. The written notifications were delivered to US consignees by traceable courier service and non-US consignees by traceable email or facsimile. The consignee notification: Advises them of the nature of the issue and the potential for an adverse patient consequence. Advises them to review their inventory and how to determine if they have affected battery chargers. Advises them to immediately stop the use or distribution of any affected product lot numbers they may have. Requests they identify and report to Integra if they do or do not have any unexpired affected product lots. Advises them replacements battery chargers will be shipped for any affected products they affirm are in their possession and how to return their affected product. Questions regarding these instructions, please contact Integra Customer Service at 1-800-431-1123 or your Integra sales
- Quantity
- 1250 battery chargers
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Canada, Australia, Arab Emirates, Denmark, Germany, Great Britain, India, Ireland, Israel, Japan, Kuwait, Lebanon, Malta., Netherlands, Panama, Poland, Saudi Arabia, Singapore, South Africa, Spain, Sweden, and Tunisia.
- Lot, serial or model codes
- Lot Numbers: IE123610, IE123710, IE123810, IE123910, IE124010, IE124110, IE124210, IE124310, IE124410, IE124510, IE124610, IE124710, IE124810, IE124910, IE125010, IE125110, IE125210, IE130110, IE130210, IE130310, IE130410, IE130510, IE130610, IE130710, IE130810, IE130910, IE131010
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HPP
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28