QUIDEL CARDIOVASCULAR INC: Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.

Recall number
Z-1974-2023
Recalling firm
QUIDEL CARDIOVASCULAR INC
Product
Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.
Status
Open, Classified
Initiated
2023-05-25
Posted
2023-07-07
Terminated
not on file
Reason
Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.
Root cause
Nonconforming Material/Component
Action
On 5/25/2023, correction notices were sent to customers and distributors who were advised the following: 1) If you have an alternate method, please discard all unused material. 2) If you do not have an alternate method, please follow these steps, as applicable, to minimize patient risk. a) Flag all negative results reported to clinicians as possibly inaccurate until lots of unaffected product are used. b) Use results from an alternate clinical laboratory analyzer when troponin results are below or close to the cutoff and myocardial infarction is suspected. c) Perform serial sampling. Keep patients until at least 3 negative troponin values have been obtained. d) Use all Triage troponin results in conjunction with the patient s risk factors, clinical presentation, EKG, and other imaging. e) Consider recommendations by the ACC, ESC guidelines and the Fourth Universal Definition of Myocardial Infarction for monitoring a patient for a rise or fall pattern of troponin. 3) If you are experiencing issues with Proficiency testing, contact your local Technical Solutions Center. 4) Complete and return the confirmation of receipt form. 5) forward this notification if the product was distributed outside of your facility. If you have further questions, please contact: - For North America, Canada, Asia-Pacific, and Latin America, please call 858.552.1100 - For Europe, Middle East, and Africa, please call +353 (91) 412 474 - For China, please call 0400 920 9366 or +86 021 3217 8300 - EU authorized representative: MDSS GmbH, Tel.: +49-511-62628630, vigilance@mdss.com On 7/12/23, Update/Reminder Notices were sent informing customers to: Discontinue use of all affected devices. Use an unaffected lot number, obtain unaffected product, use an alternate method or send patients to an alternate testing site utilizing an alternate method. If an alternate method is not available, see above.
Quantity
16,194
Distribution
US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD. OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL
Lot, serial or model codes
REF/UDI-DI/Lots(Expiration): 97000HS/30014613337911/T13666N(2023-06-26), T13667N(2023-07-03), T13669N(2023-07-08), T13706N(2023-07-21), T13826N(2023-09-03), T13944N(2023-10-14), T13948N(2023-10-20), T13949N(2023-10-22), T14019N(2023-11-05), T14020N(2023-11-05), T14023N(2023-11-11); 97021HS/30014613337935/T13665RN(2023-06-25), T13707RN(2023-06-28), T13709RN(2023-08-11), T13827RN(2023-09-04), T13829RN(2023-09-18), T13942RN(2023-10-09), T13950RN(2023-10-23); 97022HS/30014613337942/T13664RN(2023-06-20), T13708RN(2023-08-03), T13943RN(2023-10-10); 97000HSEU/30014613337928/T13668RBN(2023-07-05), T13705RN(2023-07-14), T13765RBN(2023-08-26), T13825RBN(2023-08-27), T13828RBN(2023-09-18), T13946RBN(2023-10-16); 97000HSEUJP/T13668RN(2023-07-05), T13765RN(2023-08-26), T13828RN(2023-09-18), T13946RN(2023-10-16); 97000HZ01/T13763(2023-08-19), T13764(2023-08-21), T13830(2023-09-24), T13831(2023-09-26), T13940(2023-10-01), T13941(2023-10-03), T13947(2023-10-21), T13951(2023-10-27), T14021(2023-11-06), T14022(2023-11-10), T14024(2023-11-15), T14025(2023-11-19), T14041(2023-12-13), T14042(2023-12-18), T14043(2023-12-19), T14044(2023-12-20), T14045(2023-12-24); 97000QIL/T13825RNQ(2023-08-27). 97000HS/97021HS/97022HS:Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24). 97000HS/97000HSEU/97000QIL/97021HS: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN, T13831NQ, T13831RBNQ, T13831RNQ(2023-09-26); T14021N, T14021RBN, T14021RN, T14021NQ, T14021RBNQ, T14021RNQ(2023-11-06); T14022N, T14022RBN, T14022RN, T14022NQ, T14022RBNQ, T14022RNQ(2023-11-10); T14024N, T14024RNB, T14024RN, T14024NQ, T14024RNBQ, T14024RNQ(2023-11-15); T14025N, T14025RBN, T14025RN, T14025NQ, T14025RBNQ, T14025RNQ(2023-11-19); T14041N, T14041RBN, T14041RN, T14041NQ, T14041RBNQ, T14041RNQ(2023-12-13); T14042N, T14042RBN, T14042RN, T14042NQ, T14042RBNQ, T14042RNQ(2023-12-18); T14043N, T14043RBN, T14043RN, T14043NQ, T14043RBNQ, T14043RNQ(2023-12-19); T14044N, T14044RBN, T14044RN, T14044NQ, T14044RBNQ, T14044RNQ(2023-12-20); T14045N, T14045RBN, T14045RN, T14045NQ, T14045RBNQ, T14045RNQ(2023-12-24). 97000HSEU/97000HSEUJP: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24).
510(k) numbers
["K030286"]
PMA numbers
not on file
Product code
MMI
Firm FEI
3013982035
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register