Skip to the content

FDA Medical Device Recalls (openFDA)

McKesson Provider Technologies - Medical Imaging Group: Horizon Cardiology Hemo Component Schiller Argus Pro PB-1000 Monitor unit. Argus Pro PB 1000 firmware versions 2.41+M NIBP and 2.42.01+M. A programmable diagnostic computer intended for acquiring, digitizing, storing, displaying and reviewing hemodynamic data for use in hospital cardiac catheterization laboratories. The system is cleared for us with a component physiological data collection unit - the Schiller Argus Pro PB 1000 Monitor unit.

Recall number
Z-1974-2010
Recalling firm
McKesson Provider Technologies - Medical Imaging Group
Product
Horizon Cardiology Hemo Component Schiller Argus Pro PB-1000 Monitor unit. Argus Pro PB 1000 firmware versions 2.41+M NIBP and 2.42.01+M. A programmable diagnostic computer intended for acquiring, digitizing, storing, displaying and reviewing hemodynamic data for use in hospital cardiac catheterization laboratories. The system is cleared for us with a component physiological data collection unit - the Schiller Argus Pro PB 1000 Monitor unit.
Status
Terminated
Initiated
2010-04-15
Posted
2010-07-09
Terminated
2010-07-21
Reason
McKesson has recently discovered that certain Horizon Cardiology Hemo systems were running specific Schiller Argus-Pro PB 1000 firmware versions that might cause a delay in the display of physiological parameters on the Horizon Cariology Hemo Real Time Monitor Screen.
Root cause
Software change control
Action
The firm, McKesson, sent an "Advisory Notice" letter dated April 15, 2010, to all customers. The letter describes the problem, products affected, and actions to be taken by customers. The customers were instructed not to use the Emergency Replacement Hemo kit until they receive a replacement Schiller Argus Pro PB-1000 and they have contacted McKesson Support department at 1-866-777-0202. If you have any questions regarding this notice, please call our Customer Support department at 1-0866-777-0202 and ask to speak to your support Manager.
Quantity
26 units; 19 in operational use, 8 as emergency replacement units.
Distribution
Nationwide distribution: CA, CO, MD, NH, NT, NY, TN, and TX .
Lot, serial or model codes
Argus Pro PB 1000 firmware versions 2.41+M NIBP and 2.42.01+M Serial numbers prefaced with PB-1000: 765.001471, 760.001566, 760.001793, 760.001882, 760.001920, 760.001934, 760.001970, 760.002043, 760.002113, 765.001168, 765.001307, 765.001520, 765.001553, 765.001599, 765.001609, 765.001510, 765.001632, 765.00165, 765.003014, 760.001320, 760,001552, 760.001555, 760,001920, 760.002107, 765.001224, 765.001306, 765.001608.
510(k) numbers
["K050561"]
PMA numbers
not on file
Product code
DQK
Firm FEI
3005157178
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register