AVID Medical, Inc.: Halyard CATH LAB kit. Model Number: SACL75AM.

Recall number
Z-1973-2026
Recalling firm
AVID Medical, Inc.
Product
Halyard CATH LAB kit. Model Number: SACL75AM.
Status
Open, Classified
Initiated
2026-03-13
Posted
2026-05-01
Terminated
not on file
Reason
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Root cause
Under Investigation by firm
Action
Owens and Minor/Avid Medical notified consignees on about 03/13/2026 and then sent an updated letter on about 03/18/2026 with additional kits added to the affected product list. The notice explained the issue was initiated by Medline Industries, risk associated with syringe disconnection and loose connection, and instructed consignees to identify, segregate and quarantine all affected product, and to apply warning labels to affected AVID kits to ensure all users are aware of the actions requested by Medline. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure. All usage guidelines and the Instructions for Use are provided in Medline s recall letter. Consignees were also instructed to complete and return the Recall Response Form as well as to notify customers if affected kits were further distributed to ensure all end-users are appropriately notified and maintain records of effectiveness.
Quantity
228 kits
Distribution
US distribution to FL, IL, MO, MS, NC, NE, TX.
Lot, serial or model codes
Model Number: SACL75AM. UDI-DI (Kit Lot Numbers): 10809160413586 (1599684, 1601393).
510(k) numbers
not on file
PMA numbers
not on file
Product code
OES
Firm FEI
1047429
Firm city
Toano
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register