BD SWITZERLAND SARL: BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241-0500

Recall number
Z-1973-2025
Recalling firm
BD SWITZERLAND SARL
Product
BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241-0500
Status
Open, Classified
Initiated
2025-05-12
Posted
2025-06-12
Terminated
not on file
Reason
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.
Root cause
Process control
Action
On 5/12/2025, removal notices were mailed or emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Nurse Managers, and Distributors who were asked to do the following: 1. Immediately review your inventory for affected product. Destroy all unused affected products subject to the recall following your institution s process for destruction. 2. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 3. Complete the Customer Response Form via email to bdrc53@bd.com 4. In addition, distributors were asked to identify all customers that purchased any affected product and to provide a copy of the customer letter to all customers and to advise them of the removal notification. Report any complaints experienced with the use of this product to the firm's North American Regional Complaint Center: Phone: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT, Email: productcomplaints@bd.com
Quantity
1,655,300
Distribution
Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.
Lot, serial or model codes
REF/UDI-DI/Lot(Expiration): 10012241-0500/10885403221705/23045145(April 3, 2026), 23045146(April 3, 2026), 23055091(May 4, 2026), 23065090(June 3, 2026), 23075107(July 5, 2026), 23095124(September 8, 2026), 23105027(October 3, 2026), 23115047(November 2, 2026), 23125146(December 5, 2026), 24015245(January 11, 2027), 24025887(February 7, 2027), 24026421(February 27, 2027), 24036126(March 25, 2027), 24045690(April 24, 2027), 24066893(June 25, 2027), 24075111(July 26, 2027), 24095556(September 24, 2027), 24125014(December 2, 2027), 23035049(March 1, 2026)
510(k) numbers
["K053049"]
PMA numbers
not on file
Product code
FPA
Firm FEI
3012798826
Firm city
Eysins
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register