Edwards Lifesciences, LLC: Edwards EndoReturn Arterial Cannula, model ER21B and ER23B Product Usage: The ThruPort EndoReturn arterial cannula kit (with or without hemostasis valve), an introducer, a guidewire,and a connector hub. The ThruPort catheter introducer sheath kit includes a wire reinforced cannula with hemostasis valve, an introducer, and a guidewire. The cannulae have a wire reinforced section to provide kink resistance and flexibility, tapered tips to aid in insertion and advancement into the femoral artery, a non-reinforced clamp site, a 3/8 in. (9.5 mm) barbed connector, and a hemostasis valve (available on some versions) that allows passage of catheters, such as the EndoClamp aortic catheter. The introducers accept a .038 in. (0.97 mm) guidewire and are marked to simplify assembly and indicate alignment. On some versions, a connector hub secures and immobilizes the introducer within the cannula for easier, one person insertion of the cannula/introducer assembly.

Recall number
Z-1972-2013
Recalling firm
Edwards Lifesciences, LLC
Product
Edwards EndoReturn Arterial Cannula, model ER21B and ER23B Product Usage: The ThruPort EndoReturn arterial cannula kit (with or without hemostasis valve), an introducer, a guidewire,and a connector hub. The ThruPort catheter introducer sheath kit includes a wire reinforced cannula with hemostasis valve, an introducer, and a guidewire. The cannulae have a wire reinforced section to provide kink resistance and flexibility, tapered tips to aid in insertion and advancement into the femoral artery, a non-reinforced clamp site, a 3/8 in. (9.5 mm) barbed connector, and a hemostasis valve (available on some versions) that allows passage of catheters, such as the EndoClamp aortic catheter. The introducers accept a .038 in. (0.97 mm) guidewire and are marked to simplify assembly and indicate alignment. On some versions, a connector hub secures and immobilizes the introducer within the cannula for easier, one person insertion of the cannula/introducer assembly.
Status
Terminated
Initiated
2013-07-16
Posted
2013-08-14
Terminated
2016-11-30
Reason
Edwards Lifesciences is recalling certain lots of the EndoReturn Arterial Cannula that were built with an incorrect component. This component can create a performance problem with the IntraClude IntraAortic Occlusion device.
Root cause
Nonconforming Material/Component
Action
Edwards sent an Urgent Field Safety Notice letter dated July 16, 2013 via Fed-Ex. The letter identified the affected product, problem and action to be taken. The letter included a Customer Acknowledgement Form to be completed and faxed to 800-422-9329 Attn: Recall Coordinator. If the customer does not send back the acknowledgement form, two additional contacts are made to notify the customer and request return of affected product. This is tracked by Edwards customer service and reported each week. Customer were requested to assess their inventory to assure that all affected product is identified and returned. Returned products will be destroyed. For questions call Edwards Customer Service at 800-424-3278.
Quantity
855
Distribution
US Nationwide including Canada and VA sites.
Lot, serial or model codes
Model ER21B Lots: 59496984, 59516933, 59516936, 59531385, 59531386 , 59543708   Model ER23B Lots: 59464327, 59484349, 59484350, 59484351, 59512146, 59513272, 59516940, 59534430, 59535739, 59541763
510(k) numbers
["K971291"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1713910
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register