Skip to the content

FDA Medical Device Recalls (openFDA)

Vital Signs Devices, a GE Healthcare Company: enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.

Recall number
Z-1972-2012
Recalling firm
Vital Signs Devices, a GE Healthcare Company
Product
enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.
Status
Terminated
Initiated
2012-02-20
Posted
2012-07-09
Terminated
2013-08-19
Reason
The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.
Root cause
Device Design
Action
GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included customer instructions and a Customer Response Form to be completed and returned to the firm. Contact Customer Service at 1-800-932-0760 for questions regarding this recall.
Quantity
11803 cases
Distribution
Nationwide Distribution.
Lot, serial or model codes
a) 510(k): K060537 b) Medical device listing number: D071942  All enFlow Warmer Straps in the field present a potential for tissue injury. Sold separately as Product #: 980304 and 980304EU Included as an accessory in following enFlow products:  Military Domestic 91000153 980100 91000154 980100EU 91000156 980105VS 91000103
510(k) numbers
["K060537"]
PMA numbers
not on file
Product code
LGZ
Firm FEI
2242551
Firm city
Totowa
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register