Skip to the content

FDA Medical Device Recalls (openFDA)

Starion Instruments: Starion Instruments Real Hand ThermaSeal Thermal Ligating Shears, electrosurgical cutting and coagulation device and accessories, manufactured by Starion Instruments, Sunnyvale, CA Intended use: Medical Device used for cutting and cauterization of soft tissue during surgery, and of natural or synthetic, non-metallic sutures during surgery.

Recall number
Z-1972-2010
Recalling firm
Starion Instruments
Product
Starion Instruments Real Hand ThermaSeal Thermal Ligating Shears, electrosurgical cutting and coagulation device and accessories, manufactured by Starion Instruments, Sunnyvale, CA Intended use: Medical Device used for cutting and cauterization of soft tissue during surgery, and of natural or synthetic, non-metallic sutures during surgery.
Status
Terminated
Initiated
2008-06-26
Posted
2010-07-09
Terminated
2010-10-27
Reason
Due to an improper fixture revision, device may be missing a screw at the jaw assembly, resulting in jaw components falling off during use.
Root cause
Process control
Action
The firm submitted an MDR but determined that there was no risk to health. E-mails were sent on June 26, 2008, to consignees, describing the product, the problem, and action to be taken by consignees. The consignees were ask to provide a list of all unused ThermaSeal devices and their locations, to return the products to Starion from the lots noted and to provide a copy of the attached letter to their sales force and customers if they choose. If you have any questions, please call (408) 522-5200 ext. 326 or (408) 350-9934.
Quantity
217 units
Distribution
Worldwide distribution.
Lot, serial or model codes
Model Number N-10700, lot numbers 80419, 806009, 804037, 804035, 804036, 804019A.
510(k) numbers
["K062257"]
PMA numbers
not on file
Product code
GEI
Firm FEI
1000137656
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register