Siemens Medical Solutions USA, Inc: LANTIS oncology Information System, an accessory to the Linear Accelerator, as part of the COHERENCE RT Therapist, but also cleared under the ONCOR Avante-Garde and Expression, as well as the syngo RT Therapist system, as part of the ARTISTE And PRIMUS system. Manufactured by IMPAC Medical Inc., Sunnyvale, CA, device distributed by Siemens Medical Solutions Oncology Care System, Concord, CA Electronic record management system as part of a linear accelerator system used to keep track of notes and database for oncology electronic medical records.

Recall number
Z-1970-2011
Recalling firm
Siemens Medical Solutions USA, Inc
Product
LANTIS oncology Information System, an accessory to the Linear Accelerator, as part of the COHERENCE RT Therapist, but also cleared under the ONCOR Avante-Garde and Expression, as well as the syngo RT Therapist system, as part of the ARTISTE And PRIMUS system. Manufactured by IMPAC Medical Inc., Sunnyvale, CA, device distributed by Siemens Medical Solutions Oncology Care System, Concord, CA Electronic record management system as part of a linear accelerator system used to keep track of notes and database for oncology electronic medical records.
Status
Terminated
Initiated
2009-08-21
Posted
2011-04-13
Terminated
2011-04-13
Reason
When editing the note type of a previously entered note, then reverting to the original type, data may be lost.
Root cause
Software design
Action
Siemens sent an Urgent Medical Device Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to configure their system to lock all new notes automatically. To automatically lock new notes, open eChart Navigator, right mouse click in the notes section, select "configuration" and select "Automatically Lock New Notes." If the customer chooses not to configure their system to automatically lock new notes, follow at least one of the suggestions listed below: If the customer must edit a note type, use the append feature and document that they are making an edit to the note. Lock all new notes that are entered via eChart Notes. To lock a note manually, after creating the note, right mouse click and select "Lock," Access and enter patient notes via eChart Notes until a LANTIS upgrade has been performed at your site. When entering patient notes via eChart Navigator, make sure that you are entering the notes in the correct section.\ When entering patient notes via eChart Notes, make sure that the correct note type has been selected. Customers are to keep the Customer Safety Advisory Notice together with their LANTIS 8.30 User Manual where it should remain until the measures defined above are performed. For questions call 1-888-826-9702.
Quantity
30 affected units
Distribution
Worldwide Distribution - USA including DE, WI, TX, TN, NJ, PA, NE, MA, and AL and the countries of South Africa, Australia, Ireland, Germany, Brazil, South Korea, India, and Japan.
Lot, serial or model codes
versions 8.3, CD# 9, 10 and 11, part numbers 8149259, 8164787, 8166527
510(k) numbers
["K031764","K060226","K072485","K993425"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register