FDA Medical Device Recalls (openFDA)
Bio-Rad Laboratories, Inc., Hercules, CA: Bio Rad brand D-10 Hemoglobin A1c Program Reorder Pack, which contains Update Kit D-10 A1c Program Diskette; Model Numbers: 220-0101 (Reorder Pack); 220-0115 (Program Diskette); Distributed and manufactured by: Bio-Rad Laboratories, Inc., Hercules, CA System Overview: The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, A1 c) in whole blood. The separation is based on the principles of high performance liquid chromatography. Technology Overview: The D-10 Hemoglobin Testing System is a single module designed for hemoglobin analysis in mid-volume laboratories. The D-10 is a fully integrated system for sample dilution, processing and analysis of hemoglobin, designed for use with specific Bio-Rad reagent kits. The D-1 0 incorporates the use of a dedicated software system for instrument control, data collection and analysis. The D-10 Software performs reduction of raw data collected from each analysis. Reagent Kit Information: The D-10 Hemoglobin A1c Program utilizes principles of ion-exchange High-performance liquid chromatography (HPLC). The D-10 Hemoglobin Ale Program is based on chromatographic separation of HbA1c on a cation exchange cartridge. The Bio-Rad D-10 Hemoglobin Ale Program is intended for the percent determination of hemoglobin A1 c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The Bio-Rad D-10 Hemoglobin A1c Program is intended for Professional Use Only. The D-10 Hemoglobin A1c Program is for use only with the Bio-Rad D-10 Hemoglobin Testing System.
- Recall number
- Z-1969-2011
- Recalling firm
- Bio-Rad Laboratories, Inc., Hercules, CA
- Product
- Bio Rad brand D-10 Hemoglobin A1c Program Reorder Pack, which contains Update Kit D-10 A1c Program Diskette; Model Numbers: 220-0101 (Reorder Pack); 220-0115 (Program Diskette); Distributed and manufactured by: Bio-Rad Laboratories, Inc., Hercules, CA System Overview: The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, A1 c) in whole blood. The separation is based on the principles of high performance liquid chromatography. Technology Overview: The D-10 Hemoglobin Testing System is a single module designed for hemoglobin analysis in mid-volume laboratories. The D-10 is a fully integrated system for sample dilution, processing and analysis of hemoglobin, designed for use with specific Bio-Rad reagent kits. The D-1 0 incorporates the use of a dedicated software system for instrument control, data collection and analysis. The D-10 Software performs reduction of raw data collected from each analysis. Reagent Kit Information: The D-10 Hemoglobin A1c Program utilizes principles of ion-exchange High-performance liquid chromatography (HPLC). The D-10 Hemoglobin Ale Program is based on chromatographic separation of HbA1c on a cation exchange cartridge. The Bio-Rad D-10 Hemoglobin Ale Program is intended for the percent determination of hemoglobin A1 c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The Bio-Rad D-10 Hemoglobin A1c Program is intended for Professional Use Only. The D-10 Hemoglobin A1c Program is for use only with the Bio-Rad D-10 Hemoglobin Testing System.
- Status
- Terminated
- Initiated
- 2010-09-08
- Posted
- 2011-04-13
- Terminated
- 2011-06-29
- Reason
- The firm discovered that the Update Kit Floppy Diskette included in the reorder pack was mislabeled with an incorrect Lot Number for the Calibrator/Diluent Set as: S01819; The correct Lot Number for the Calibrator/Diluent Set is: S01704.
- Root cause
- Other
- Action
- The Bio-Rad Laboratories, Inc, sent a Medical Device Correction letter dated 9/8/10, to all affected customers via Federal Express addressing this mislabeling issue, a new Update Kit Floppy Diskette, Lot No. BA02210 has been produced, and Bio-Rad requests that you please destroy the Lot No. AA01815. If you have any questions please contact your regional Bio-Rad office. You may also call (510) 724-7000.
- Quantity
- not on file
- Distribution
- Nationwide Distribution including : PA, OH, MO, CA, TX, FL, MA, MD, KY, OH, MI, NJ, and WA.
- Lot, serial or model codes
- Lot 20001814; Exp. February 28,2012
- 510(k) numbers
- ["K031043"]
- PMA numbers
- not on file
- Product code
- LCP
- Firm FEI
- 2915274
- Firm city
- Hercules
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28