FDA Medical Device Recalls (openFDA)
Datascope Corporation: Sensation Plus 7.5Fr. 40cc Intra-Aortic Balloon Catheter with accessories, Part Number 0684-00-0568-01 Product Usage: The Sensation Plus 7.5Fr. 40cc Intra-Aortic Balloon Catheter with accessories is a device that includes a catheter, an insertion kit and two STATLOCK IAB stabilization devices. It is a cardiac assist device intended to provide counterpulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.
- Recall number
- Z-1966-2019
- Recalling firm
- Datascope Corporation
- Product
- Sensation Plus 7.5Fr. 40cc Intra-Aortic Balloon Catheter with accessories, Part Number 0684-00-0568-01 Product Usage: The Sensation Plus 7.5Fr. 40cc Intra-Aortic Balloon Catheter with accessories is a device that includes a catheter, an insertion kit and two STATLOCK IAB stabilization devices. It is a cardiac assist device intended to provide counterpulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.
- Status
- Terminated
- Initiated
- 2019-05-28
- Posted
- not on file
- Terminated
- 2023-09-25
- Reason
- The device was distributed with the outer carton labeled with the manufacture date and the expiry date reversed. This showed the product as expired before it was manufactured.
- Root cause
- Employee error
- Action
- An Urgent Medical Device Recall notification letters dated 5/28/19 was sent to the customer. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached URGENT MEDICAL DEVICE RECALL CUSTOMER RESPONSE FORM on page 3, to acknowledge that you have received the Sensation Plus 7.5 Fr. 40cc Intra-Aortic Balloon Catheter with Accessories URGENT MEDICAL DEVICE RECALL. Please fax or email the completed Customer Response Form to 1-585-486-5605.
- Quantity
- 471
- Distribution
- The products were distributed to the following US states: FL. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Estonia, France, Germany, Ireland, Italy, Kuwait, Lithuania, Malaysia, Mexico, Netherlands, Poland, Russia, Saudi Arabia, Slovakia, Sweden, Switzerland, United Kingdom, Ukraine.
- Lot, serial or model codes
- UDI Code: 10607567108063; Lot Codes: 3000082472, 3000084069, 3000085231, 3000086994, 3000087688, 3000091205, 3000091206, 3000091828
- 510(k) numbers
- ["K122628"]
- PMA numbers
- not on file
- Product code
- DSP
- Firm FEI
- 1000222374
- Firm city
- Fairfield
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28