Integra LifeSciences Corp.: MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws

Recall number
Z-1961-2014
Recalling firm
Integra LifeSciences Corp.
Product
MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws
Status
Terminated
Initiated
2014-06-09
Posted
2014-07-01
Terminated
2014-11-06
Reason
Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.
Root cause
Under Investigation by firm
Action
A customer notification letter was sent on 6/9/14 to their customers and distributors who purchased the Integra NewPort System MIS Rod, 2 to inform them there may be a potential that the hex on some NewPort System MIS Rod, 2 are up to 30 degrees offset from specification requirements causing misalignment with the inserter. Customers are instructed to complete and return the customer response form and return by email or fax as indicated on the form. Customers with questions are instructed to contact Customer Service at (866) 942-8698.
Quantity
94 units
Distribution
Worldwide Distribution. US including states of CA, FL, IL, MD, MO, TN, TX, and countries of Australia, Czech Republic and Spain.
Lot, serial or model codes
Lot No.  Z21781A (45mm NewPort MIS Rod, 2) Z21956A (50mm NewPort MIS Rod, 2) Z21782A (55mm NewPort MIS Rod, 2) Z21957A (60mm NewPort MIS Rod, 2) Z21784A (75mm NewPort MIS Rod, 2) Z21958A (80mm NewPort MIS Rod, 2)
510(k) numbers
["K083089"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register