B. Braun Interventional Systems: Celsite Access Port Model ST305P. The Celsite Access Port model ST305P is a member of the Celsite product line of implantable port and catheter systems.

Recall number
Z-1961-2009
Recalling firm
B. Braun Interventional Systems
Product
Celsite Access Port Model ST305P. The Celsite Access Port model ST305P is a member of the Celsite product line of implantable port and catheter systems.
Status
Terminated
Initiated
2009-06-22
Posted
2009-09-03
Terminated
2012-06-05
Reason
B. Braun Interventional Systems was notified by B. Braun Medical France, the manufacturer of the Celsite Access port that they had received 11 reports of polyurethane (type P) catheter disconnections. B. Braun Medical France has decided to recall all Celsite ports supplied with the affected catheter lot. To date, B. Braun Interventional Systems (US) has not received any reports of catheter dis
Root cause
Nonconforming Material/Component
Action
B. Braun Interventional Systems, Inc. issued an "Urgent: Medical Device Recall" notice dated June 22, 2009. The letter identified the problem and product. The letter reminded the consignee of the general instruction for use of access ports and instructed them to contact patients who had the Celsite port system implanted. For further questions, contact B. Braun Interventional Systems, Inc. at 1-800-523-9676 extension 4818.
Quantity
40 in US
Distribution
Nationwide Distribution -- NY.
Lot, serial or model codes
Lot Numbers: H3238310 and I0059950.
510(k) numbers
["K061424","K952548"]
PMA numbers
not on file
Product code
LJT
Firm FEI
3006332832
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register