Skip to the content

FDA Medical Device Recalls (openFDA)

Spine Wave, Inc.: Spine Wave Access Kit, Catalog Number 10-1530 Package contents: RLSP494-Rhythmlink Monopolar Stimulating Probe and 10-1525 Disposable Dilator Set. Product Usage: The Spine Wave Access Kit is comprised of a sterile monopolar stimulating probe (individually packaged by Rhythmlink International) and a set of three sterile disposable dilators (individually packaged by Spine Wave). The Rhythmlink International Monopolar Stimulating Instrument is indicated for stimulation of cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots.

Recall number
Z-1960-2015
Recalling firm
Spine Wave, Inc.
Product
Spine Wave Access Kit, Catalog Number 10-1530 Package contents: RLSP494-Rhythmlink Monopolar Stimulating Probe and 10-1525 Disposable Dilator Set. Product Usage: The Spine Wave Access Kit is comprised of a sterile monopolar stimulating probe (individually packaged by Rhythmlink International) and a set of three sterile disposable dilators (individually packaged by Spine Wave). The Rhythmlink International Monopolar Stimulating Instrument is indicated for stimulation of cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots.
Status
Terminated
Initiated
2015-05-04
Posted
2015-07-02
Terminated
2017-01-24
Reason
Mislabeled outer carton: Monopolar Probe is listed as 200 x 1.5mm when it should be listed as 280 x 1.5mm
Root cause
Packaging change control
Action
Spine Wave sent an Urgent: Medical Device Recall letter dated June 11, 2015 to affected customers. The letter identified the affect product, problem and actions to be taken. Customers were instructed to acknowledge receipt of notification by returning the enclosed Recall Notice Acknowledgement Form in the stamped envelope provided. For questions, contact Spine Wave directly at 203-944-9494 and ask for the Recall Coordinator or by email at rsmith@spinewave.com.
Quantity
121 units
Distribution
US Nationwide Distribution in the state of CA, FL, GA, LA, MA, MD, MI, MO, NC, NH, OH, OR, PA, TX, and VA.
Lot, serial or model codes
Lot Codes: 092514Z, 120814A, 011415H and 022415F
510(k) numbers
["K072736"]
PMA numbers
not on file
Product code
ETN
Firm FEI
3004638600
Firm city
Shelton
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register