Canon Medical System, USA, INC.: Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is a general purpose x-ray system that employs Solid State Imager(s), SSXI, which converts x-rays directly into electrical signals which can, after appropriate processing be displayed on LCD monitors or printed to a medical grade image printer. The system console is PC based devise that allows for worklist management, image storage, image processing, image exporting and image printing. The system may be equipped with a table and/or vertical wall unit, is configurable with up to two x-ray tubes, and has an auto switching function. This system is intended for use in conjunction with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck and limbs. The system is used for image acquisition, image display and transmission/output or images to external devices.

Recall number
Z-1958-2019
Recalling firm
Canon Medical System, USA, INC.
Product
Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is a general purpose x-ray system that employs Solid State Imager(s), SSXI, which converts x-rays directly into electrical signals which can, after appropriate processing be displayed on LCD monitors or printed to a medical grade image printer. The system console is PC based devise that allows for worklist management, image storage, image processing, image exporting and image printing. The system may be equipped with a table and/or vertical wall unit, is configurable with up to two x-ray tubes, and has an auto switching function. This system is intended for use in conjunction with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck and limbs. The system is used for image acquisition, image display and transmission/output or images to external devices.
Status
Terminated
Initiated
2018-10-16
Posted
not on file
Terminated
2020-07-20
Reason
software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.
Root cause
Software design
Action
On October 16, 2019, the firm sent an urgent medical device correction letter via USPS, return receipt to all customers who were distributed the affected device. The notification letter ask customers to do the following: 1. Should the problem occur before the modified software is installed on your system, please reduce the number of other devices which use the same wireless channel as this system as much as possible to avoid heavy wireless traffic. 2. Please share this information with all users and reviewing radiologist as well as clinical engineering and the biomedical group at your facility 3. Please complete and return the customer reply form via email to RAffairs@us.medical.canon or by fax to regulatory affairs at 877-349-3054. 4. If you have any questions regarding the letter, please contact Terry Schultz, Regulatory Affairs Manager at (800) 421-1968 or contact your local Canon Representative at (800) 521-1968.
Quantity
46
Distribution
US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL
Lot, serial or model codes
No./Serial No./Manufacture Date:  No 1./U8A1312001/JANUARY 2013;  No.2/U8A1312002/JANUARY 2013;  No.3/U8A1342003/APRIL 2013;  No.4/U8A1362004 /JUNE 2013;  No. 5 U8B13X2006/OCTOBER 2013; No. 6 U8B13Z2008/DECEMBER 2013; No. 7 U8B1412009/JANUARY 2014; No. 8 U8B1432010 /MARCH 2014; No. 9 U8B1442011/APRIL 2014; No.10 U8B1462013/JUNE 2014; No.11 U8B1472014/JULY 2014; No.12 U8B1482015/AUGUST 2014; No.13 U8B14Y2016/NOVEMBER 2014; No.14 U8B14Z2017/DECEMBER 2014; No.15 U8B14Z2018/DECEMBER 2014; No.16 U8B14Z2019/DECEMBER 2014; No.17 U8B1512020/JANUARY 2015; No.18 U8B1522021/FEBRUARY 2015; No.19 U8B1522022/FEBRUARY 2015; No.20 U8B1522023/FEBRUARY 2015; No.21 U8B1522024/FEBRUARY 2015; No.22 U8B1612025/JANUARY 2016; No.23 U8B1632026/MARCH 2016; No.24 U8B1642027/APRIL 2016; No.25 U8B1642028/APRIL 2016; No.26 U8B1642029/APRIL 2016; No.27 U8B1662030 /JUNE 2016; No.28 U8B1677001/JULY 2016; No.29 U8B16X2032/OCTOBER 2016; No.30 U8B1722033 /FEBRUARY 2017; No.31 U8B1752034/MAY 2017; No.32 U8B1762035/JUNE 2017; No.33 U8B1782036 /AUGUST 2017; No.34 U8B17Z2037 /DECEMBER 2017; No.35 U8C1812038/JANUARY 2018; No.36 USA1462001/JUNE 2014; No.37 USA1492002/SEPTEMBER 2014; No.38 USA14Z2003/DECEMBER 2014; No.39 USA1512004/JANUARY 2015; No.40 USA1532005/MARCH 2015; No.41 USA15Z2006/DECEMBER 2015; No.42 USA1662007/JUNE 2016; No.43 USA1682008/AUGUST 2016; No.44 USA16X2009/OCTOBER 2016; No.45 USA16Y2011/NOVEMBER 2016; No.46 U8A1372005/JULY 2013; No.47 U8B13X2007/OCTOBER 2013; No.48 USA16X2010/OCTOBER 2016
510(k) numbers
K122842
PMA numbers
not on file
Product code
MQB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source