FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Cios Alpha; The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients
- Recall number
- Z-1958-2015
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Cios Alpha; The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients
- Status
- Terminated
- Initiated
- 2015-05-05
- Posted
- 2015-07-02
- Terminated
- 2017-01-13
- Reason
- under certain circumstances the Cios Alpha system may freeze during a procedure.
- Root cause
- Software design
- Action
- Siemens sent an Safety Advisory Notice dated May 5 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were alerting them about the issue and risks involved. Customers were provided work-around instructions until a software fix is released. For further questions please call (610) 219-6300.
- Quantity
- 37
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Model Number of device 10308191, serial numbers: 11032 11044 11005 10115 11056 10057 10076 10019 10021 10035 10037 10073 11074 11075 11077 11083 10078 11066 11067 11013 10063 10082 11094 10089 11080 11009 11018 11082 10058 10098 11015 11016 11038 10096 10116 11035 11064
- 510(k) numbers
- ["K132094"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28