Skip to the content

FDA Medical Device Recalls (openFDA)

Dental Equipment LLC: Dental chairs distributed under several brand names such as DCI Equipment; Marus MaxStar; Marus NuStar; and Pelton & Crane. Part number for the bolt: 022R065 Product is labeled in part: DCI Equipment, Mfg By: Dental Equipment, LLC. DC1235, Marus Dental, Mfg By: Dental Equipment, LLC. DC1490, DC1690, and DC1700, and DC1702, Pelton & Crane, Spirit SP17.

Recall number
Z-1958-2011
Recalling firm
Dental Equipment LLC
Product
Dental chairs distributed under several brand names such as DCI Equipment; Marus MaxStar; Marus NuStar; and Pelton & Crane. Part number for the bolt: 022R065 Product is labeled in part: DCI Equipment, Mfg By: Dental Equipment, LLC. DC1235, Marus Dental, Mfg By: Dental Equipment, LLC. DC1490, DC1690, and DC1700, and DC1702, Pelton & Crane, Spirit SP17.
Status
Terminated
Initiated
2011-03-01
Posted
2011-04-12
Terminated
2011-09-21
Reason
Link arm assembly that mounts the rear mounted accessories to the dental chair may potentially break and fall off the dental chair.
Root cause
Process control
Action
On 03/08/11, Dental Equipment LLC. started sending out the Medical Device Recall notification letter via USPS 1st class mail to US consignees and Federal Express mail to foreign consignees. The firm also sent a recall notification letter to consignees who only purchased dental accessories such as Rear Mounted Cuspidor with Coverset, Arm Articulating Rear Mounted Cuspidor - Marus Gray, Maxstar Orbit Cuspidor utility W/ Porcelain bowl, and zinc plated bolts. The letter identified the affected dental chair as Marus brand with the following model numbers: DC1490, DC1690, DC1700 and DC1702. The firm's corrective action including the replacement of the affected link arm assembly with a new link arm assembly. The firm's field representative will contact each consignee for an appointment to replace the affected link arm assembly. Consignees can contact the Dental Equipment LLC at 503-537-3617 for any questions.
Quantity
55 chairs and 11 replacement bolts distributed to US and foreign consignees outside the US.
Distribution
Worldwide distribution to USA, Australia, Canada, Ireland, South Africa, and Vietnam.
Lot, serial or model codes
Part number for the bolt: 022R065  Chair Serial Numbers: 1961, 2127, 2260, 239446, 239446, 240619, 241535, 242384, 242711, 242862, 244489, 244502, 244530, 244977, 245874, 246275, 246526, 247557, 247813, 247976, 248040, 248871, 249015, 249235, 249238, 249441, 249784, 218756C, 2250, 2251, 240126, 240126A , 240605, 240605A, 240605B, 240605C, 240605D, 240605E, 241379, 241379A, 241379B, 242154J, 242154K, 242154L, 245767, 245767F, 246769, 246769A , 247199AE, 247199AJ, 248129F, 248321, 248321A, 248378C, 248824, and 248824A.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NRU
Firm FEI
1000523896
Firm city
Newberg
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register