Skip to the content

FDA Medical Device Recalls (openFDA)

Microbiologics Inc: KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC 17960; 0621K, Staphylococcus aureus derived from NCTC 12493

Recall number
Z-1957-2025
Recalling firm
Microbiologics Inc
Product
KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC 17960; 0621K, Staphylococcus aureus derived from NCTC 12493
Status
Open, Classified
Initiated
2025-05-27
Posted
2025-06-11
Terminated
not on file
Reason
Affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.
Root cause
Mixed-up of materials/components
Action
On May 27, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were informed of the affected products containing the wrong microorganisms. Customers were instructed to review lab procedures to understand how this information affects their usage, and then to use or discard the product depending on the lab procedures and how the issue affects usage. Distributors of the affected product were instructed to contact end users who have received the affected items.
Quantity
3 total
Distribution
US Nationwide distribution in the states of California and Pennsylvania and the country of New Zealand.
Lot, serial or model codes
01065K: UDI-DI 30845357032868, lot 1065-25-51; 0621K: UDI-DI 30845357014666, lot 621-73-21
510(k) numbers
["K861022"]
PMA numbers
not on file
Product code
JTR
Firm FEI
2150138
Firm city
Saint Cloud
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register